液体流率MR测量实验研究 被引量:2

Experimental Study of Quantitative Flow Measurement by Using MRI
下载PDF
导出
摘要 目的探讨稳定流动状态下磁共振信号强度与液体流速关系,为横窦血流量测定提供适当方法。材料和方法在实验水模上放置内径3.4mm塑胶管,用高压注射器以不同的流量向管道内注入水,流量的变化范围从0.1~1.6ml/s。用1.5T超导磁共振仪单个无门控2DPC序列进行扫描,对液体磁共振信号强度和管道内实际流速采用相关回归分析方法进行统计分析。结果磁共振信号强度(y)和液体速度(x)呈明显线性关系,y=68.914x+357.206,R2=0.998。结论相位对比法是一种有效的流量测量方法。 Purpose:To find the relation between the signal intensity of MR and the flow velocity under steady flow condition in order to develop a suitable protocol for using this method to measure volumetric flow rate in the transverse sinus system.Materials and Methods:A phantom consisted of a tube3.4mm in internal diameter laid on fluid-filled phantom and water as the circulating fluid.Steady flow was generated by an electric high power injector in the velocity rang of0.1~1.6ml/s.A superconduct MR imager operating at1.5T was used for in vitro study.MR imaging was performed by means of single ungated two dimensional phase contrast technique(2D PC),and the linear regression analysis was used for statistic analysis beˉtween the signal intensities determined by MR and the true velocities in the tube phantom.Results:A straight-line correlation was produced between signal intensity(y)determined with PC MR and the true velocity(x)in the tube:y=68.914x+357.206,R 2 =0.998.Conclusion:Phase contrast method is an efˉfective method for measuring volumetric flow rate.
出处 《中国医学计算机成像杂志》 CSCD 2004年第3期145-148,共4页 Chinese Computed Medical Imaging
关键词 液体流率 MR测量 磁共振信号 颅内压 血液流速 Magnetic resonance imaging Phantoms Flow quantification
  • 相关文献

参考文献9

  • 1Nayler DL, Firmin DN, Longmore DB. Blood flow imaging by cine magnetic resonance. J Comput Assist Tomogra 1986; 10:715 - 722
  • 2Meier D, Maier S, Bosiger P. Quantitative flow measurements on phantoms and on blood vessels with MR. Magn Reson Med 1988;8:25 - 34
  • 3Duerk JL, Pattany PM. In - plane flow velocity quantification along the phase encoding axis in MRI. Magn Reson Imaging 1988;6:321 - 333
  • 4Kondo C, Caputo GR, Semelka R, et al. Right and left ventricular stroke volume measurements with velocity - encoded cine MR imaging:in vitro and in vivo validation. A JR 1991;157:9 - 16
  • 5Heinrich M, Robert RE, Klaus UW, et al. Middle cerebral artery:determination of flow velocities with MR angiography. Radiology 1991; 181:527 - 530
  • 6Spritzer CE, Pelc NJ, Lee JN, et al. Rapid MR imaging of blood flow with a phase - sensitive, limited - flip - angle, gradient recalled pulse sequence: preliminary experience. Radiology 1990;176:255 - 262
  • 7Lorie R, Norbert J, Stephen C, et al. Arterial and venous blood flow: nonivaeive quantitation with MR imaging. Radiology 1992;185:809 - 812
  • 8Peeters F, Luypaert R, Eisendrath H, et al. Time resolved flow quantification with MRI using phase methods: a linear systems approach. Magn Reson Med 1995;33:337- 354
  • 9Bernstein MA, Ikezaki Y. Comparison of phase - difference and complex - difference processing in phase - contrast MR angiography. Magn Reson Imaging 1991; 10: 725- 729

同被引文献19

  • 1李彩英,田建明,刘辉,陆建平,黄勃源,刘怀军.利用磁共振相位对比法对非搏动流体模型的定量测量研究[J].中国医学影像技术,2006,22(4):535-537. 被引量:10
  • 2de Haan MW,Kouwenhoven M,Kessels AG,et al.Renal artery blood flow:quantification with breath-hold or respiratory triggered phase-contrast MR imaging[J].Eur Radiol,2000,10(7):1133-1137.
  • 3Koskenvuo JW,Hartiala JJ,Knuuti J,et al.Assessing coronary sinus blood flow in patients with coronary artery disease:a comparison of phase-contrast MR imaging with positron emission tomography[J].AJR,2001,177(5):1161-1166.
  • 4Yildiz H,Erdogan C,Yalcin R,et al.Evaluation of communication between intracranial arachnoid cysts and cisterns with phase-contrast cine MR imaging[J].AJNR,2005,26(1):145-151.
  • 5Oktar SO,Yucel C,Karaosmanoglu D,et al.Blood-flow volume quantification in internal carotid and vertebral arteries:comparison of 3 different ultrasound techniques with phase-contrast MR imaging[J].AJNR,2006,27(2):363-369.
  • 6Pujadas S,Reddy GP,Weber O,et al.Phase contrast MR imaging to measure changes in collateral blood flow after stenting of recurrent aortic coarctation:initial experience[J].J Magn Reson Imaging,2006,24(1):72-76.
  • 7Sankari S,Lehmann P,Gondry JC,et al.Phase-contrast MR imaging support for the diagnosis of aqueductal stenosis[J].AJNR,2009,30(1):209-214.
  • 8Pelc NJ,Herfkens RJ,Shimakawa A,et al.Phase contrast cine magnetic resonance imaging[J].Magn Reson Q,1991,7(4):229-254.
  • 9Greil G,Geva T,Maier SE,et al.Effect of acquisition parameters on the accuracy of velocity encoded cine magnetic resonance imaging blood flow measurements[J].J Magn Reson Imaging,2002,15(1):47-54.
  • 10Lethimonnier F,Bouligand B,Thouveny F,et al.Error assessment due to coronary stents in flow-encoded phase contrast MR angiography:a phantom study[J].J Magn Reson Imaging,1999,10(5):899-902.

引证文献2

二级引证文献15

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部