期刊文献+

自制颈椎融合器的实验研究 被引量:1

Experimental study of the self-made cervical fusion cage
原文传递
导出
摘要 目的 探讨自制颈椎融合器在动物体内的应用 ,为进一步临床应用提供实验依据。方法 应用实验动物山羊 2 0头 ,分为实验组和对照组 ,分别测试抗压缩力、椎间隙高度及组织反应。结果  (1)抗压缩力 :自制颈椎融合器 (3 5 8.64± 15 .63 )N ,自体骨植入组为 (2 68.82± 11.3 6)N ,差异有显著性 (P <0 .0 5 ) ;(2 )椎间高度 :自制颈椎融合器组为 (12 .0 5± 0 .16)mm ,自体髂骨植入组为 (12 .0 5± 0 .14 )mm ,差异无显著性 (P >0 .0 5 ) ;(3 )自制颈椎融合器有较好的生物相容性。结论 自制颈椎融合器强度大、生物相容性好、稳定性好 ,是一种比较理想的颈椎融合器。 Objective To investigate the application of the self-made cervical vertebral fusion cage in goats to provide experimental data for future clinical usage.Methods Twemtu goats were divided into two groups:control group and experimental group.The strength of anti-compression,intervertebral space height and tissue reactions were tested in both groups.Results There was significant difference in the strength of anti-compression between the groups of self-designed cervical vertebral fusion cage (358.64±11.36) N and the autograft bone (268.82±11.36) N ( P <0.05).There was no significant difference between the experiment group (12.05±0.16) mm and the control group (12.08±0.14) mm in the intervertebral space height ( P >0.05).The self-made cervical vertebral fusion cage has good biocompatibility.Conclusion The self-made cervical vertebral fusion cage is capable of high strength,good biocompatibility and stability.
出处 《中华实验外科杂志》 CAS CSCD 北大核心 2004年第5期611-612,共2页 Chinese Journal of Experimental Surgery
基金 河南省自然科学基金资助项目 (2 0 0 0 4 0 2 4 1 0 0 )
关键词 颈椎融合器 实验研究 山羊 高分子聚乙烯 颈椎病 Cervical vertebral fusion cage Goat Macromolecular polyethylene
  • 相关文献

参考文献3

二级参考文献5

  • 1赵定麟,脊柱外科学,1996年,294页
  • 2Wang J C,Spine,1998年,23卷,2450页
  • 3任先军,中华创伤杂志,1998年,14卷,增刊,49页
  • 4Mei F R,Spine,1997年,2083页
  • 5任先军,梅芳瑞,孙玉华,周军海.颈椎前融合保留终板的临床研究[J].中华创伤杂志,1998,14(S1):55-57. 被引量:8

共引文献50

同被引文献5

  • 1郭永飞,陈德玉,徐建伟,王新伟,王良意,何志敏,陈宇,袁文,贾连顺.颈前路钛网植骨融合术后钛网沉陷的原因分析[J].中国脊柱脊髓杂志,2005,15(7):409-413. 被引量:43
  • 2Juan SU,Jaypal RS,Edward AM,et al.Comparison between anterior cervical discectomy fusion and cervical corpectomy fusion using titanium cages for reconstruction:analysis of outcome and long-term followup.European Spine Journal,2009,18:654-662.
  • 3Wilke H J,Kettler A,Claes LE.Are sheep spines a valid biomechanical model for human spines? Spine,1997,22:2365-2374.
  • 4Kandziora F,Pflugmacher R,Scholz M,et al.Comparison between sheep and human cervical spines:an anatomic,radiographic,bone mineral density,and biomechanical study.Spine,2001,26:1028-1037.
  • 5王良意,陈德玉,郭永飞,王新伟,徐建伟,王以进.颈前路可调式融合固定器的研制及生物力学测试[J].中国矫形外科杂志,2004,12(11):854-857. 被引量:12

引证文献1

二级引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部