摘要 本文描述了一种新的测量二氧化硫的荧光型光纤传感器。溶于内充液的奎宁被用作敏感物,通过气透膜将内充液与水溶液分开。传感器对二氧化硫的响应范围是3×10^(-6)~3×10^(-4)mol/L。报道了传感器的重现性、稳定性、响应时间和选择性。用该传感器测定了含二氧化硫的实际样品并与碘量法进行了比较。 A new fluorescence fiber optic sensor for sulfur dioxide measurement has been described. Quinine dissolving in internal solution was used as a sensitive material. The internal solution layer was separated from the aqueous solution by a gas-permeable membrane. The response range of the sensor for sulfur dioxide is from 3×10-4 mol/L down to 3×10-6 mol/L. Reproducibility, stability, resp-ponse time and selectivity of the new fiber optic sensor are reported. With this sensor, practical samples containing sulfur dioxide were measured. The results are compared to measurments with iodime-try.
出处 《高等学校化学学报》 SCIE EI CAS CSCD 北大核心 1993年第4期476-479,共4页 Chemical Journal of Chinese Universities
基金 国家教育委员会优秀年轻教师基金 湖南省科委基金
关键词 光纤传感器 荧光测定法 二氧化硫 Gas sensor, Sulfur dioxide fiber optic sensor,Fluorimetry
  • 相关文献

参考文献6

  • 1陈国珍,荧光分析法(第2版),1990年
  • 2张志琪,化学学报,1987年,45卷,497页
  • 3虞振新,离子选择电极分析应用指南,1986年
  • 4朱元保,高等学校化学学报,1983年,4卷,571页
  • 5团体著者,污染源统一监测分析方法.废气部分,1983年
  • 6团体著者,分析化学,1981年