期刊文献+

流动注射分析血清中葡萄糖的电化学检测池 被引量:1

An Electrochemical Detector Cell for Flow Injection Analysis of Glucose in Serum
下载PDF
导出
摘要 测定血清中葡萄糖含量是临床检验的重要项目之一。何立千曾对各种血糖测定方法做过综述。葡萄糖氧化酶-4-氨基安替比林-苯酚光度法(酶制剂法)虽能较好地测定血清中葡萄糖的含量,但成本较高,耗时长。因此有必要研究一种快速、廉价、稳定的分析方法。人们考虑到酶的专属性,对葡萄糖氧化酶电极做了大量工作,然而,这种电极的寿命仍不理想。 An electrochemical detector cell for FIA of glucose in serum has been developed. It was a three electrodes system involved a homemade copper-PTFE composite plated film electrode (working) and other two electrodes, silver (reference) and platinum(auxiliary) wires. The carrier liquid for FIA was 0. 5 mol/L NaOH and flowed through the cell at 2.1 mL/min. The working electrode potential was set at 0. 6 V(vs. reference) by potentiostat. Sampling 30 ftL test solution which was 1% serum sample diluted with water, a peak current was recorded by chronome-try. The peak current was linearly related with the concentration of glucose in the range of 5 × 10-3-2×10-1 mmol/L. The analytical results from our method were pretty correlated to those from glucose oxidase-4-aminoanti-pyrine-phenol method(Trinder method). The regression equation from the results of twenty samples was Y=-0. 039 + 1. 005X, r=0. 9999. It was shown that our method was speedy, cheap and reliable.
机构地区 吉林大学化学系
出处 《高等学校化学学报》 SCIE EI CAS CSCD 北大核心 1993年第11期1526-1527,共2页 Chemical Journal of Chinese Universities
基金 国家自然科学基金
关键词 血清 葡萄糖 流动注射分析 Glucose in serum, Flow injection analysis, Electrochemical detector cell
  • 相关文献

参考文献2

  • 1许宏鼎,1993年
  • 2何立千,中华医学检验杂志,1982年,5卷,1期,49页

同被引文献6

引证文献1

相关作者

内容加载中请稍等...