牛奶成分近红外光谱分析中系统漂移的补偿 被引量:4

Compensation of System Drift in Compositional Analysis of Milk by Using Near-infrared Spectroscopy
下载PDF
导出
摘要 为实现牛奶中蛋白质、脂肪等成分含量的快速无损检测,本文论述了采用近红外漫反射光谱(1000~1700nm)分析牛奶主要成分含量的基本方法,定量分析了仪器信噪比与系统测量误差之间的传递关系。在此基础上,进一步探讨了采用双光路设计以及参考液光谱自动校正方法消除系统漂移的可行性。实验结果表明:采用参考液光谱自动校正方法可以有效地克服系统漂移,校正后脂肪、蛋白质的实际预测相对偏差(RMSEP)与参考值平均值之比分别降低了0.14%和0.72%。 A method to realize rapid nondestructive examination to the composition of milk by using Near-infrared Scattered reflection spectroscopy(1000~1700nm)was explored,and quantitative relation of signal-to-noise(SNR)and measurement error of the system were analyzed.The feasibility of double light path compensation method and reference solution spectral auto-correction method against system drift during building model and measuring process were respectively researched and experimentally verified.It was concluded that the reference solution spectral auto-correction method could effectively compensate system drift, and the predictive error in practice(ratio of Root Mean Square Error of Prediction(PMSEP)to mean of reference)of fat and protein reduce correspondingly 0.14% and 0.72% after correction with the method,which was useful for improving long-time stability of the measuring system.
出处 《激光与红外》 CAS CSCD 北大核心 2004年第3期209-212,共4页 Laser & Infrared
关键词 光谱分析 漫反射 双光路设计 自动校正 系统漂移 spectral analysis scattered reflection double light design auto-correction system drift
  • 相关文献

参考文献3

  • 1Lu Wanzhen,Yuan Hongfu,Xu Guangtong.Modern Near Infrared Spectral Analysis Technology[A].Beijing:China Petrochemical Press,2000,8-10.
  • 2Yongdong Wang,David J Veltkamp,Bruce R Kowalski.Multivariate instrument standardization[J].Analytical Chemistry,1991,63(23):2750-2756.
  • 3Chad E Anderson,John H Kailivas.Fundamentals of calibration transfer through Procrustes analysis[J].Applied Spectroscopy,1999,53(10):1268-1275.

同被引文献93

引证文献4

二级引证文献44

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部