期刊文献+

自聚焦多元辐射器超声热场的计算机模拟与实验研究 被引量:3

Computer Simulation and Experimental Research of Multi-element Ultrasonic Applicator
下载PDF
导出
摘要 分析了非相干自聚焦多元辐射器的声场结构,对该声场进行了模拟计算,绘制了声强分布图,并利用自行研制的九元阵装置进行了实验研究。结果表明:非相干自聚焦多元辐射器可以产生具有较好聚焦特性的超声热场;焦域吸收率SAR大于80%,焦域温度在2s内达45°C,焦域直径在 5mm以内。上述特性适合于高强度超声聚焦热疗使用。 The structure of the self-focused and non-interferential transducer array in their ultrasonic field has been analyzed in detail in this paper.At the same time, with the help of the fast computing ability of the computer, the simulation analysis has been done, and the ultrasonic field distribution is drawn in figures. The results show that: the applicator has a good focus character in the deep tissue region. Specific absorption rate (SAR) >80%, the focus region can reach the temperature of higher than 45°C only used less than 3 seconds, the diameter is within 5mm, which met to the need of high strength ultrasound hyperthermia.
出处 《压电与声光》 CSCD 北大核心 2004年第4期325-327,330,共4页 Piezoelectrics & Acoustooptics
基金 天津市自然科学基金资助项目(003609111)
关键词 超声热疗 热场分布 自聚焦阵 ultrasound hyperthermia heat field distribution self-focused transducer array
  • 相关文献

参考文献5

二级参考文献25

  • 1林世寅,胡自省,毛慧生,张天泽.第七届国际肿瘤热疗会议概述[J].中国肿瘤临床,1996,23(9):671-675. 被引量:11
  • 2王鸿樟.换能器及聚焦系统[M].上海:上海交通大学出版社,1996..
  • 3Hill,C.R.[J].Ultrasound Med.Biol.,1994;20:271~277.
  • 4Frizzell,L.A.[J].IEEE Trans Ultrason.Ferroelect.Freq.Control,1998;35:578~581.
  • 5Lizzi,F.L,Ostromogilsky,M.[J].Ultrasound Med.Biol.,1987;13:607~618.
  • 6Hill,C.R,Rivens,I,Vaughan,M.G.et al.[J]Ultrasound Med.Biol.,1994;20:259~269.
  • 7Robison,T.C,Lele,P.P.[J].J.Acoust.Soc.Am,1972;51:1333~1351.
  • 8Chen.L,Rivens,I,ter Haar,G.R.et al.[J]Ultrasound Med.Biol.,1993;19:67~74.
  • 9Yang R,Reilly C R,Rescorla FJ et al.[J].Pediatr Surg,1992;27:2,246~250;discussion 250~251.
  • 10Watkin,N.A,ter Haar,G.R,Morris,S.B.et al.[J].Br J.Urol 1995;75(suppl l),1~8.