高效液相色谱仪器系统对红霉素肟分析的影响 被引量:1

Effect of High Performance Liquid Chromatographic Instrument System on the Analysis of Erythromycin A Oxime
下载PDF
导出
摘要 探讨了日立HITACHIL 710 0高效液相色谱 (HPLC)、岛津LC 6A型液相色谱、WatersTM4 74液相色谱系统及相关色谱柱对红霉素肟及其相关物质的色谱条件影响 ,实验表明 ,在其他色谱条件相对稳定的条件下 ,不同型号仪器和色谱柱对样品分离度及保留时间有较大影响 ,摸索到了不同仪器系统下的稳定分析条件 ,取得了相对良好的线性相关性 ,对不同环境下进行红霉素肟的质量检测具有指导意义。 A HPLC chromatography for the determination of erythromycin A oxime and relative compounds was studied, amd the effect of chromatography systems including a HITACHI L-7100, a Shimadzu LC-6A, a Waters(TM) 474 and relative columns was analyzed. it was revealed that different HPLC apparatus and columns have obvious impact on the peak separation and retention time under the general chromatographic condition. The suitable chromatographic conditions for several different chromatography systems were summarized with good linear relationship, which is very significant to the quality control of erythromycin A oxime and relative compounds.
出处 《光谱学与光谱分析》 SCIE EI CAS CSCD 北大核心 2004年第9期1148-1150,共3页 Spectroscopy and Spectral Analysis
基金 国家经贸委技术创新项目 (0 1BK 0 0 9)资助
关键词 高效液相色谱仪器系统 红霉素肟 质量检测 红霉素 药物中间体 HPLC chromatographic instrument erythromycin A oxime
  • 相关文献

参考文献6

  • 1[1]Clark R F, Ma Z K, Wang S Y, Griesgraber G, Tufano M, Yong H, Li L P, Zhang X L, Nilius A M, Chu D T W. Bioorg. Medi.1 Chem. Lett., 2000, 10: 815.
  • 2[3]Morimoto S, Adachi T, Matsunaga T, Kashimura M, Watanabe Y, Sota K. U. S. Patent, 6342590 B1, 2002.
  • 3[5]Robert R. Willkening. USP 5912331, 1999.
  • 4[6]Harang V, Westerlund D. Chromatographia, 1999, 50(9-10): 525.
  • 5[7]Moheb M N, Terry J T. J. Liq. Chrom. rel. Technal., 1997 20(4): 553.
  • 6[8]Chieko T, Hisakazu O, Erika H, Hajime K, Hitoshi S, Tatsuji I. J. Chrom. B, 2000, 783: 405.