期刊文献+

^(103)Pd支架预防兔髂动脉支架后再狭窄的实验研究

Experimental study of the preventive effect of^(103)Pd stent on in-stent restenosis in rabbit iliac artery
下载PDF
导出
摘要 目的 探讨^(103)Pd支架置入对兔髂动脉支架后再狭窄的预防作用。方法 20只新西兰大白兔按支架放射性活度分为6组[2.22(n=3),5.55(n=4),9.25(n=4),14.8(n=3),22.2(n=3),33.3 MBq(n=3)],于每只动物一侧髂动脉内置入不同活度的^(103)Pd支架,对侧髂动脉内置入非放射性支架作对照。高脂饲养28d后进行髂动脉血管造影和定量组织病理学分析,观察治疗效果。结果组织病理学定量分析示不同剂量^(103)Pd支架组新生内膜面积和狭窄百分比均明显低于对照组,但其效应不呈明显剂量依赖性。结论 ^(103)Pd支架能明显抑制支架置入后兔髂动脉新生内膜增生和支架后再狭窄的发生。 Objective To investigate the ability of γ-emitting 103Pd radioactive stent implantation for preventing in-stent restenosis in iliac artery of rabbits. Methods Twenty New Zealand white rabbits were divided into six groups on the basis of radioactivity loading [2.22 (n = 3), 5.55 (n = 4), 9.25 (n = 4), 14.8 (n = 3), 22.2 (n = 30), 33.3 MBq (n = 3)].103Pd stent was implanted in one of the two iliac arteries of each rabbit and nonradioactive stent in the contralateral artery. The rabbits were fed with high fat diet. Treatment efficiency was assessed by angiography and quantitative histomorphometry of the stented iliac segments 28 d after the implantation. Results Quantitative histomorphometry analysis indicated that103 Pd stents induced a significant reduction in neointimal area and the percentage of stenosis within radioactive stents was lower compared with that in the control stents. Nevertheless, the suppression of neointima formation by103 Pd stent implantation was not in a dose-dependent manner. Conclusions Low-dose intravascular brachytherapy via a γ-particle-emitting 103 Pd radioactive stent potently inhibits neointimal hyperplasia and prevents the in-stent restenosis in rabbit iliac artery.103 Pd stent implantation probably can be employed as a novel means to prevent in-stent restenosis.
出处 《中华核医学杂志》 CAS CSCD 北大核心 2003年第6期331-333,T002,共4页 Chinese Journal of Nuclear Medicine
关键词 ^103Pd支架 预防作用 髂动脉支架后再狭窄 放射疗法 Iliac artery Stents Palladium Hyperplasia Radiotherapy Rabbits
  • 相关文献

参考文献12

  • 1李勤.不同剂量测量体系的剂量学结果比较[J].中华放射医学与防护杂志,2000,20(3):201-202. 被引量:1
  • 2Fajardo LF. The nature of arterial restenosis after angioplasty. Int J Radiat Oncol Biol Phys, 1998, 40: 761-763.
  • 3Fischman DL, Leon MB, Baim D, et al. A randomized comparison of coronary stent placement and balloon angioplasty in the treatment of coronary artery disease. New Engl J Med, 1994, 331: 496-501.
  • 4Carter AJ, Laird JR, Kafe WM, et al. Coronary stenting with a novel stainless steel balloon expendable stent: determinants of neointimal formation and changes in arterial geometry after placement in an atherosclerotic model. J Am Coll Cardiol, 1996, 27: 1270-1277.
  • 5Condado JA, Waksman R, Gurdiel O, et al. Long-term angiographic and clinical outcome after percutaneous transluminal coronary angioplasty and intracoronary radiation therapy in human. Circulation, 1997, 96: 727-732.
  • 6Wilcox JN, Waxman R, King SB, et al. Role of adventitia in the arterial response to angioplasty. In: Waksman R, King SB, Mould RF, eds.Vaseular brachytherapy. Veenendal: Nucletron, 1996. 17-30.
  • 7Scott NA, Ross CE, Dunn B, et al. Identification of a potential role for the adventitia in vascular lesion formation after balloon overstretch injury of porcine coronary arteries. Circulation, 1996, 93: 2178-2187.
  • 8Browne E, Firestone RB. Table of radioisotopes. In: Shirley VS, Ed.New York: Wiley Interscience Publications, 1988. 102-105.
  • 9Brenner DJ, Amols HI. Enhanced risk from low-energy screen-film mammography X-rays. Br J Radiol, 1989, 62: 910-914.
  • 10McLemore LB, Horton JL, Antolak JA. Calculation of the 3-D dose distribution surrounding a 103Pd stent. Cardiovasc Radiat Med, 2001, 2:181-190.

二级参考文献3

  • 1JJG10 2 6-91 光子和高能电子束吸收剂量测量方法,1991年
  • 2JJG6 6 4-90 医用加速器X辐射源,1990年
  • 3JJG 589-89.60 Co 远距离治疗辐射源,1989年