光程长对光谱测量误差的影响 被引量:5

Influence of Pathlength on the Error of Spectral Measurement
下载PDF
导出
摘要 光程长的选择在光谱测量方法设计中非常重要,但实际测量中很难实现精确最佳光程长.从灵敏度的角度研究了在近红外光谱测量中,光程长和测量灵敏度之间的定量关系,讨论了它们对测量结果的影响:单位光程长吸光度越小,则光程长对测量误差的影响越小,可选择光程长范围就越宽;大于最佳光程长区域相对于光程较小区域,光程长变化对测量结果影响较小,可选光程过窄,应通过减小单位光程吸光度来扩展可选区域.对于单个样品,当最佳光程长获取存在较大误差时,应使所选光程长落在大于最佳光程长区域;当一组样品分别对应不同最佳光程长时,也应选择较大光程.在10个光程长下,通过6000cm-1~5900cm-1长波段不同浓度葡萄糖水溶液光谱测量实验,对理论结果进行验证,二者吻合较好. Selection of pathlength is key to near infrared spectral measurement.In this paper,the quantitative relationship between pathlength and measuring sensitivity is studied, with their influence on measurement result discussed.The less the absorptivity of unit pathlength is,the less the influence of pathlength on the measurement error is and the larger the available pathlength range can be.The influence of pathlength variation is weaker in the range the pathlength is longer than the optimum pathlength.If acceptable pathlength range is too narrow,it should be expanded by decreasing the absorptivity of the sample.For a single sample,the pathlength should be selected in the range that is longer than optimum pathlength if a large pathlength selection error is unavoidable.As for a group of samples which correspond to different optimum pathlengths,the longer one is preferred.The experiments of NIR spectrometry of aqueous glucose solutions among wavelength of 6 000 cm^(-1)—5 900 cm^(-1) on ten different pathlengths verify the results of above-mentioned theory.
出处 《天津大学学报(自然科学与工程技术版)》 EI CAS CSCD 北大核心 2004年第10期906-909,共4页 Journal of Tianjin University:Science and Technology
基金 国家"十五"科技攻关项目(2001BA706B 21) 国家自然科学基金资助项目(30170261).
关键词 光程长 灵敏度 光程长选择范围 测量误差 pathlength sensitivity available pathlength range measurement error
  • 相关文献

参考文献6

  • 1Jensen P S, Bak J. Near-infrared transmission spectroscopy of aqueous solutions: Influence of optical pathlength on signal-to-noise ratio[J]. Applied Spectroscopy,2002,56(12):1 600-1 606.
  • 2Willard H, Merritt L, Dean J, et al. Instrument Methods of Analysis[M]. Belmont, CA: Wadsworth,1981.
  • 3Hazen K H,Arnold M A,Small G W.Measurement of glucose in water with first-overtone near-infrared spectra[J]. Applied Spectroscopy,1998,52(12):1 597-1 603.
  • 4Xu K X, Yutaka Y,Harumie U, et al. Apparatus and Method for Optically Measuring Concentration of Components [P].USA:USP 5602647, 1997-02-11.
  • 5许禄.化学计量学方法[M].北京:科学出版社,1997.287-372.
  • 6王宏,李庆波,刘则毅,徐可欣.遗传算法在近红外无创伤人体血糖浓度测量基础研究中的应用[J].分析化学,2002,30(7):779-783. 被引量:13

二级参考文献1

  • 1许禄.化学计量学方法[M].北京:科学出版社,1997.101.

共引文献39

同被引文献34

引证文献5

二级引证文献27

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部