Measurement of Radial Compliances of Small Diameter Artificial Vascular Graft
下载PDF
导出
摘要 【目的】设计小径人工血管的径向顺应性测量的方法和装置。【方法】由激光测微器、压力传感器、A/D卡、微电脑和脉动循环回路等部件制成血管顺应性测定装置,血管外径信号由激光测微器记录,内腔压力信号由压力传感器记录,经A/D转换和微电脑处理,两者的波形和数值均在显示器上实时和同步显示。【结果】该装置的压力测量范围0~300mmHg,分辨率0.015mmHg,直径测量范围0~9mm,分辨率0.005mm,对各种尺寸和性质的人工血管的径向顺应性的测量结果表明,能满足使用要求。【结论】该装置和方法可测量内径2~4mm,顺应性大于1.0×10-2·%/mmHg的各种规格小径人工血管的径向顺应性,是小径人工血管性能研究的一种工具。 To design a method and a device to measure radial compliances of small diameter artificial vascular graft.A device was made by a laser micromete r,a pressure transducer,a A /D card,a computer,and a pulsed circulation loop.The size signals of grafts external diameter were picked by the laser micrometer,the pressure sign als were picked by the pressure transducer and transforme d by A /D card and the computer to wave f orms and digital data,that were displayed synchronously o n the monitor of computer in real time .For the device diameter measured was in the range of0~9mm and the resolution was in 0.005mm,the pressure measured was in the range of 0~300mmH g and the resolution was in 0.015mmHg .Small diameter artificial vascular grafts with different dimension and property were measured.The results showed that the device could meet demands for mea suring radial compliance of those grafts.[Conclusion]This device can measure artificial vascu lar grafts with internal diameter in2~4mm and radial compliance in more than 1.0×10 -2 ·%/mmHg.It is a new tool for studying p roperty of artificial vascular graft.
出处 《中山大学学报(医学科学版)》 CAS CSCD 北大核心 2004年第5期404-407,412,共5页 Journal of Sun Yat-Sen University:Medical Sciences
基金 国家自然科学基金资助项目(30070223)
关键词 人工血管 顺应性 激光 处理 微电脑 记录 方法 装置 性能研究 循环回路 small diameter artificial vascular graft radial compliance laser micrometer pressure transducer
  • 相关文献

参考文献6

  • 1Zidi M, Cheref M. Mechanical analysis of a prototype of small diameter vascular prosthesis: numerical simulations [J]. Comput Biol Med, 2003,33(1): 65-75.
  • 2Hsu S H, Helen K. On matching compliance between canine carotid arteries and polyurethane graft [J] . Artificial Organes, 1997,21(12): 1247-54.
  • 3Zhang Z, King M W, Guidoin R, et al. Morphological physical and chemical evaluation of the vascugraft arterial prosthesis: comparison of a novel polyurethane device with other microporous structures [J] . Biomaterials, 1994,15(7): 483-501.
  • 4Martz H, Beaudoin G, Paynter R. et al. Physicochemi cal characterization of a hydrophilic microporous polyurethane vascular graft[J]. J Biomed Meter Res, 1987, 21(4): 399-412.
  • 5潘仕荣,廖凯荣,易武,郑欢玲,王琴梅.聚六亚甲基碳酸酯聚氨酯脲的合成研究[J].中山大学学报(自然科学版),2003,42(4):34-37. 被引量:9
  • 6Seifalian A M, Salacinski H J, Tiwari A. et al, in vivo biostability of a poly(earbonate-urea) urethane graft[J] .Biomaterials, 2003, 24(14): 2549-57.

二级参考文献4

共引文献8

同被引文献7

引证文献1

二级引证文献3

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部