三维超声心动图评价右室容量的实验研究 被引量:1

The experimental study of right ventricular volume evaluated by three-dimensional echocardiography
下载PDF
导出
摘要 目的 评价三维超声重建定量测定右心室容量的准确性。方法  15个离体新鲜猪心模型 ,采用三维超声测量其右室容量 ,并与实测值比较。结果 新鲜猪心模型右室容量的三维超声测量值与实测值分别为 2 3 5 8± 6 4 2ml和 2 1 70± 5 5 1ml,两者间无统计学差异 (P >0 0 5 ) ;且高度相关 (r=0 93,P <0 0 0 1)。结论 三维超声心动图为定量测定右室容量提供了准确和切合临床实用的新方法。用Pearson相关分析。结果一、右室容量 :三维法测量值与实测值分别为2 3 5 8± 6 4 2和 2 1 70± 5 5 1ml ,两者间无统计学差异 (P >0 0 5 ) ,详见表 1。表 1 三维法测量离体新鲜猪心右室容量与实测值比较编号三维法 (ml)实测值 (ml)编号三维法 (ml)实测值 (ml)115 92 16 8917 2 615 32 2 2 3 12 0 5 10 2 6 5 82 2 23 18 2 0 17 5 1118 2 82 1 64 2 6 10 2 2 612 18 5 416 95 3 8 83 3 2 113 2 6 85 2 3 762 5 65 2 3 2 14 2 1 9619 773 4 2 93 5 2 15 19 5 417 882 3 462 0 4二、三维法测量值与实测值之间关系 :相关分析表明 ,两者之间高度相关 (r=0 93,P <0 0 0 1) ,其直线回归方程为Y =1 0 9X +0 0 6 ,(见图 1)。图 1 三维法测量值与实测值之间的关系讨论目前 ,多种影像学技术如心血管造影、 Objective To assess the accuracy of three-dimensional echocardiography in the quantitative measurement of RV volume by using excised fresh pig heart model.Methods In 15 excised fresh pig heart model,three-dimensional echocardiography was used to reconstruct the right ventricular cavity,and compared with actual volume.Results It was shown that the RV volume measured by three-dimensional echocardiography agreed well with actual volume(r=0.93,P<0.001),and there is no significant difference between the mean values(23.58±6.42 vs 21.70±5.51 ml,P>0.05).Conclusion It is concluded that three-dimensional echocardiography supplies a convenient and accurate method for quantitative assessment for RV volume.
出处 《江苏医药》 CAS CSCD 北大核心 2004年第10期738-739,共2页 Jiangsu Medical Journal
关键词 三维超声心动图 实验研究 定量测定 猪心 三维超声重建 右心室 离体 Echocardiography,three-dimensional Ventricular volume,right
  • 相关文献

参考文献6

  • 1Yao JF, Kasprzak JD, Nosir YFM, et al. Appropriate 3-dimensional echocardiography data acquisition interval for left ventricular volume quantification: implications for clinical application. J Am Soc Echocardiogr, 1999,12:1053-1057.
  • 2Heusch A,Rubo J,Krogmann ON,et al. Volume measurement of left ventricle in children with congenital heart defects: 3-dimensional echocardiography versus angiocardiography. Cardiology,1999,92:45-52.
  • 3Linker DT, Moritz MW, Pealman AS, et al. A new three-dimensional echocardiographic method of right ventricular volume measurement: in vitro validation. J Am Coll Cardiol, 1986,8:101-106.
  • 4Schindera ST,Mehwald PS,Sahn DJ,et al. Accuracy of real-time three-dimensional echocardiography for quantifying right ventric ular volume: static and pulsatile flow studies in an anatomic in vitro model. J Ultrasound Med, 2002,21:1069-1075.
  • 5Hubka M, Bolson EL, McDonald JA, et al. Therr-dimensional echocardiographic measurement of left and right ventricular mass and volume: in vitro validation. Int J Cardiovasc Imaging, 2002,18:111-118.
  • 6丁建东.动态三维超声心动图的临床应用[J].心血管病学进展,2002,23(6):342-345. 被引量:6

二级参考文献1

共引文献5

同被引文献12

  • 1Vogel M, Gutberlet M, Dittrich S, et al. Comparison of transthoracic three dimensional echocardiography with magnetic imaging in assessment of right ventricular volume and mass[J]. Heart, 1997, 78(2): 127.
  • 2Veldtman G R, Razack V, Siu S, et al. Right ventricular form and function after percutaneous atrial septal defect device closure[J]. J Am Coll Cardiol, 2001, 37(8): 2108.
  • 3Kort H W, Balzer D T, Johnson M C, et al. Resolution of right heart enlargement after closure of secundum atrial septal defect with transcatheter technique[J]. J Am Coll Cardiol, 2001, 38(5): 1528.
  • 4Muller S, Bartel T, Pachinger O, et al. 3-D echocardiography: new developments and future prospects[J]. Herz, 2002, 27(3): 227.
  • 5Sugeng L, Weinert L, Thiele K, et al. Real-time three-dimensional echocardiography using a novel matrix array transducer[J]. Echocardiography, 2003, 20: 623.
  • 6Jiang L, Levine R A, Weyman A E, et al. Echocardiographic assessment of right ventricular volume and function[J]. Echocardiography, 1997, 14: 189.
  • 7Fujimoto S, Mizuno R, Nakagawa Y, et al. Estimation of the right ventricular volume and ejection fraction by transthoracic three-dimensional echocardiography. A validation study using magnetic imaging[J]. In J card Imaging, 1998, 14(6): 385.
  • 8Yao J F, Kasprzak J D, Nosir Y F M, et al. Appropriate 3-dimensional echocardiography data acquisition interval for left ventricular volume quantification: implications for clinical application[J]. J Am Soc Echocardiogr, 1999, 12: 1053.
  • 9Godoy I E, Bednarz J, Sugeng L, et al. Three-dimensional echocardiography in adult patients: comparison between transthoracic and transesophageal reconstruction[J]. J Am Soc Echocardiogr, 1999, 12: 1045.
  • 10Heusch A, Rubo J, Krogmann ON, et al. Volumetric analysis of the right ventricle in children with congenital heart defects: comparison of biplane angiography and transthoracic 3-dimensional echocardiography[J]. Cardiol Young, 1999, 9(6): 577.

引证文献1

二级引证文献3

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部