起搏器自动夺获功能中备用脉冲发放情况的临床观察 被引量:7

Observation of Backup Safe Pulses Releasing in AFFINITY Series Pacemakers
下载PDF
导出
摘要 目的研究患者日常活动时,植入的起搏器自动夺获功能工作中备用脉冲的发放。方法对15例植入Affinity系列起搏器的患者进行24小时动态心电图记录,采取人工逐跳分析的人机对话方法对发放的备用起搏脉冲进行计数及分类。结果本组患者随访前一个月的起搏阈值0.62±0.26V~1.17±0.55V,随访当日的自动起搏电压1.06±0.29V。15例患者(100%)在真性或假性心室融合波后有备用脉冲的发放,1例(6.7%)因未检测到ER波(EvokedResponse)而发放,2例(13.3%)因自动输出的起搏电压不能夺获心室而发放,1例(6.7%)因感知不良,起搏脉冲落在自身心室除极后未夺获心室发放。没有发现自动输出电压起搏心室失败又无备用脉冲发放的情况。结论自动阈值检测及备用脉冲的发放受真性或假性心室融合波,以及快速心律失常的影响,高于起搏阈值0.25V的实际起搏电压能够有效起搏心室。 Objective To study the backup safe pulses releasing in AFFINITY Series Pacemakers during daily-life. Methods We studied 24-hour Holter recorded in 15 patients who had been implanted the Affinity SR series pacemakers(St. Jude Medical), and classified different kinds of the backup safe pulses manually by analyzing beat-by-beat. Results Backup safe pulses released after fusion/pseudo-fusion of ventricular complexes were in all 15 patients. Backup safe pulses were released because E/R was not detected in 1 patient(6.7%), and auto putout couldn't capture the ventricular muscles in 2 patients(13.3%).In 1 case(6.7%)because ventricular sensitivity was not sufficient, some pulses were released after self repolariztion and could not capture the heart, which contributed to the backup safe pulse release. We hadn't found that no backup safe pulses released when the auto putout pace pulses lost capture. Their chronic pacing threshold in the latest month was between0.62±0.26V and 1.17±0.55V,and the auto output on the testing day was 1.06±0.29V. Conclusion Auto threshold examining and backup safe pulses releasing can influenced by fusion/pseudo-fusion and tachyarrhythmia. Pacing by 0.25V higher than the threshold can capture the cardiac muscles effectively.
出处 《临床心电学杂志》 2004年第3期172-175,共4页 Journal of Clinical Electrocardiology
关键词 心室 起搏器 患者 发放 自动夺获功能 假性 随访 临床观察 真性 24小时动态心电图 Capture threshold Auto-capture Backup safe pulses
  • 相关文献

参考文献5

  • 1Suri R, Harthorne JW, Galvin J. Automatically optimizing pacing output; an excellent idea, but with potentially lethal pitfalls.Pacing Clin Electrophysiol 2001 Apr;24(4 Pt 1);520-3.
  • 2Simeon L, Duru F, Fluri M, et al, The impact of automatic threshold tracking on pulse generator longevity in patients with different chronic stimulation thresholds. Pacing Clin Electrophysiol 2000 Nov;23(11 Pt 2);1788-91.
  • 3Schuchert A,Ventura R, Meinertz T;AUTOCAP Investigators. Adjustment of the evoked response sensitivity after hospital discharge in pacemaker patients with automatic ventricular threshold tracking activated.Pacing Clin Electrophysiol; 2001 Feb;24(2);212-6.
  • 4陈新主编.临床心律失常学-电生理和治疗.北京;人民卫生出版社,2000,1824-1827.
  • 5郭继鸿,张海澄主译.Barold S,Mugica著.心脏起搏的最新进展-21世纪的目标.北京;北京医科大学出版社,2000,297-301.

同被引文献19

引证文献7

二级引证文献12

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部