摘要 氢化物原子吸收法是原子吸收分光光度法的一个重要分支.笔者应用氢化物电热原子化装置测定血清的硒具有用血量少、成本低、方法简便、精度佳等优点,适合一般医院开展科研和临床测定.采用这种技术可减少试样的基体干扰,提高了分析灵敏度,而且具有良好的线性关系,其测定检出限为1.2μg/L,浓度变异系数为6.3%。 The hydride atomic absorption method is an important branch of atomic absorption spectrophotometry. The selenium (Se) in blood serum is determined by the author with hydride electrothermal atommizing device. This method has many advantages such as less blood quantity demand. low cost. easy operation and high precision. and is suitable for scientific research and clinic determination at ordinary hospitals. This technique can adopted to reduce the matrix disturbance of specimen, increase analytical sensitivity and has good linear relation, Its detection limit is 1. 2μg/L. The coefficient of concentration variation is 6. 3%.
作者 周世兴
出处 《广东微量元素科学》 CAS 1994年第2期27-29,共3页 Trace Elements Science
关键词 热原 人血清 血量 临床 医院 氢化物原子吸收法 原子吸收分光光度法 分支 检出限 线性关系 Hydride atomic absorption method,Blood serum,Selenium.