摘要
FIA-Sub methyl bule spectrophotometry was adopted in the paper.When temperature and reagent proportion was determined,the influence of peristaltic pump rate,the length of reaction tube and injection volume were studied to optimize the analytical conditions.The result showed that the linear range,the limit of and detection the accuracy were 1~25mg/L,1mg/L,and 3.56%(RSD),respectively.The method was successfully applied to the measurement of sulfide in leather wastewater.
FIA-Sub methyl bule spectrophotometry was adopted in the paper.When temperature and reagent proportion was determined,the influence of peristaltic pump rate,the length of reaction tube and injection volume were studied to optimize the analytical conditions.The result showed that the linear range,the limit of and detection the accuracy were 1~25mg/L,1mg/L,and 3.56%(RSD),respectively.The method was successfully applied to the measurement of sulfide in leather wastewater.
出处
《化学研究与应用》
CAS
CSCD
北大核心
2004年第6期843-845,共3页
Chemical Research and Application
基金
国家高技术研究发展计划(863计划)经费资助项目(2001AA647020)