一种新型人造心脏双叶瓣跨瓣压差的研究 被引量:3

STUDY OF THE TRANSVALVULAR GRADIENT OF A NEWLY CHINA-MADE BILEAFLET HEART VALVE PROSTHESIS
下载PDF
导出
摘要 为了观察新型人造双叶瓣的跨瓣压差 ,在实验室、动物和临床上对该种人造瓣作了跨瓣压差的测定。 1.实验室测定 :在模拟人体生理循环条件下进行流体力学测试。通过自编数据处理程序对参数进行自动分析。 2 .动物实验测定 :手术中对置换 2 1mm型号二尖瓣的 6只羊用双导管法测定血流动力学 ;对 2只术后存活两年零五个月的羊先用超声法 ,然后用导管法并在多巴酚丁胺负荷下作血流动力学测定。 3.临床患者测定 :对 14只主动脉瓣和 10只二尖瓣用导管法和超声法测定。结果 :1.实验室测定 :测试瓣的ΔP ,均随心率增加而减低 ,ΔP在10mmHg以下。 2 .动物实验 :术中 6只羊的ΔP ,导管法 :5 .2mmHg± 1.7mmHg。 2只术后两年零五个月羊的ΔP ,超声法 :6 .8mmHg。 3.临床患者 :14只主动脉瓣ΔP ,导管法 :6 .2 4mmHg~ 4 .10mmHg ;超声法 :9.4 2mmHg~ 7.4 8mmHg。 10只二尖瓣ΔP ,导管法 :2 .1mmHg~ 1.9mmHg ;超声法 :5 .3mmHg~ 4 .1mmHg。结论 :新型人造双叶瓣平均跨瓣压差较低 ,具有良好的血流动力学效果。 To assess the transvalvular gradient of a newly China - made bileaflet heart valve prosthesis, the hydrodynamics were performed in the test and the hemodynamic performance were measured in the animals and patients. Methods 1. Test: the prosthesis has been tested in a pulsatile flow simulator in the aortic position. 2. Animal experiments: Six sheep that underwent mitral replacement with the prosthesis of 21mm were measured by open cardiac catheterization intraoperatively. Doppler echocardiography and open cardiac catheterization under dobutamine stress was performed in two sheep 2.5 years post - implant. 3. Patients measurements: 14 cases of aortic valve and 10 cases of mitral valve in the patients who underwent valve replacement with the heart valve prosthesis have been measured by open cardiac catheterization and Doppler echocardiography. Result: 1. Test: The newly heart valve prosthesis had the lower mean transvalvular gradient which was in the range of below 10 mmHg at any given tissue annulus diameter. 2. Animal experiments: The transvalvular gradients of 6 sheep were 5. 2mmHg ± 1.7mmHg intraoperatively and 2 sheep 2.5 years post - implant were 6.1mmHg ± 0.3mmHg by catheterization. 3. Patients measurement: The mean transvalvular gradients in the aortic position measured by means of catheterization and echocardiography were 6.26mmHg - 4.10mmHg and 9.42mmHg - 7.48 mmHg, respectively. The gradients in the mitral position were 2.10mmHg - 1.9mmHg and 5.281mmHg - 4. 10mmHg, respectively. Conclusion The results of this study demonstrated that the newly valve prosthesis has excellent hemodynamic performance.
出处 《中国生物医学工程学报》 EI CAS CSCD 北大核心 2004年第6期549-553,共5页 Chinese Journal of Biomedical Engineering
关键词 人造心脏瓣 心脏瓣膜置换术 血流动力学 Echocardiography Hemodynamics Hydrodynamics Patient monitoring Pulsatile flow Stress analysis Tissue
  • 相关文献

参考文献11

  • 1Akins CW. Mechanical cardiac valvular prostheses[J]. Ann Throrac Surg.1991,52:161-172.
  • 2Akins CW. Results with mechanical cardiac valvular prostheses[J]. Ann Thorac Surg. 1995, 60: 1836-1844.
  • 3Wilkins GT, Gillam LD, Kritzer GL, et al. Validation of continuous-wave Doppler echocardiographic measurements of mitral and tricuspid prosthetic valve gradients: A simultaneous Doppler-catheter study[J]. Circulation. 1986,74;786-795.
  • 4Labovitz AJ. Assessment of prosthetic heart valve function by Doppler echocardiography[J]. Circulation. 1989, 80:707-709.
  • 5Izzat MB, Birdi I, Wilde P, et al. Comparison of hemodynamic performance of St. Jude Medical and CarboMedics 21mm aortic prostheses by means of Dobutamine stress echocardiography[J]. J Thorac Cardiovasc Surg 1996,111:408-415.
  • 6Rashitan MY, Stevenson DA, Allen DT, et al. Flow characteristics of four commonly used mechanical heart valves[J]. Am J Cardiol 1986,58: 743-752.
  • 7Cooper DM, Stewart WJ, Schiavone WA, et al. Evaluation of normal prosthetic valve function by Doppler echocardiography[J]. Am heart J 1987,114:576.
  • 8Holen J, Simonsen S, Froysaker T. An Ultrasound Doppler technique for the noninvasive determination of the pressure gradient in Bjork-Shiley mitral valve[J].Circulation. 1979,59:436-442.
  • 9Baumgartner H, Khan S, DeRobertis M, et al. Effect of prosthetic aortic valve design on the Doppler-catheter gradient correlation: an invitro study of normal St.Jude. Medtronic-Hall, Starr-Edwards and Hancock valves[J]. J Am Coll Cardiol 1992,19:324-332.
  • 10Ihen H, Molstad P, Simonsen S, et al. Hemodynamic evaluation of the CarboMedics prosthetic heart valve in aortic position: Comparison of noninvasive and invasive techniques[J]. Am Heart J. 1992,123:151.

同被引文献55

引证文献3

二级引证文献5

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部