摘要
利用电泳中介微分析通过与邻菲罗啉的络合反应完成了对Fe2+与Cu2+的同时测定.在压差作用下,淌度较小的邻菲罗啉在金属离子区带之前首先进样,加高电压后,两区带电泳在磷酸缓冲液中(pH=4.5)混合、反应形成络合物并实现电泳分离.研究了缓冲液pH、试剂和样品区带长度对分离效果的影响.对于Fe2+与Cu2+两离子,方法的检出限(S/N=3)分别为1.6和11 μmol /L;线性范围分别为5.0×10-6~3.0×10-4 mol/L (r=0.9979)和8.0×10-5~3.0×10-4 mol/L (r=0.9996).采用标准矿样验证了测定Fe2+的可行性.
Simultaneous determination of two transition metal ions (Fe 2+ , Cu 2+ ) after complexation with o phenanthroline (phen) were investigated by electrophoretically mediated microanalysis (EMMA). The slower o phenanthroline zone was injected spatially before a faster analyte zone containing Fe 2+ and Cu 2+ in a fused silica capillary under pressure. Upon application of the high voltage, the analyte zone was electrophore tically mixed with the reagent zone, then the complexes were formed and separated in a phosphate buffer (pH=4.5). The separation behavior was investigated with respect to dependence on the buffer pH, analyte and reagent zone length. In this method, the linear calibration ranges were 5~300 ( r =0.9979) and 80~300 μmol/L ( r =0.9996) with detection limits of 1.6 and 11 μmol /L for Fe 2+ and Cu 2+ , respectively. The method was also validated by analysis of the standard reference mineral samples.
出处
《分析化学》
SCIE
EI
CAS
CSCD
北大核心
2005年第1期41-44,共4页
Chinese Journal of Analytical Chemistry
基金
东省教育厅科技计划资助项目(No. 03A06)
关键词
进样
铜(Ⅱ)
铁(Ⅱ)
FE^2+
邻菲罗啉
离子
矿样
电泳
缓冲液
试剂
Electrophoretically mediated microanalysis,iron (Ⅱ), copper (Ⅱ), o-phenanthroline, capillary electrophoresis