TJ-Ⅳ型VCG微机自动分析系统的研制与临床应用

Development and Applications of an Auto-Analyzing System for Model TJ-Ⅳ Vector-Cardiogram
下载PDF
导出
摘要 研制TJ-Ⅳ型心电向量图微机自动分析系统。与人工测算结果相比,本系统对正交心电图P波、Q波、R波、S波、R'波、S'波、T波起点和止点识别的准确性达98%,对各波谷点和峰点的识别准确性达100%。该系统开发诊断参数591项,包括46项创新指标。分析其中12项指标.人工测算结果与微机自动测算结果基本一致。该系统解决了心电向量图检测与分析中所存在的若干技术难题.提高了本系统抗干扰性能和安全性能,图形自动识别准确率高,测算参数精确度高,检测速度快。 A new computer-assisted vector-cardiogram analyzing system Model TJ-Ⅳ developed based on Model TJ-Ⅲ, has been using in the routine clinical work in order to evaluate its features and performances. The system employs a 586 computer with a CPU of 120 MHz, a special low-noise amplifier, a 12 bit A/D tranducer and the C language for programming. The examinations of 206 cases were performed and all the vector-cardiograms were analyzed by the computer system and by manipulative methods respectively. In comparison with the manipulative methods the system has a very high accuracy of picture-recognition. The accuracy for distinguishing the onsets and terminals of orthogonal ECG waves is 98% while that for distinguishing the peaks and troughs of the waves is 100%. These waves include P, Q, R, S, R' and S' waves. The new system is capable to provide the parameters of more than 591 items, including 46 newly-developed diagnostic parameters. The testing and analyzing of 12 parameters of orthogonal ECG and plane VCG have proved that the results of the aboved two methods have no difference. The new system has a very high accuracy of picture-recognition and index calculation with many technical problems existing in the old versions, solved - a great improvement of safety and anti-interference and an increase of the detecting & diagnostic speed.
出处 《中国医疗器械杂志》 CAS 2005年第1期19-22,共4页 Chinese Journal of Medical Instrumentation
基金 湖北省重点攻关项目基金资助武汉市科委晨光计划项目基金资助
关键词 心电向量图 正交心电图 微机辅助诊断 vector-cardiogram, orthogonal electrocardiogram, computer-assisted diagnosis.
  • 相关文献

参考文献10

  • 1郭小梅,王兆椿.慢性肺心病QRS向量系统性研究[J].心电学杂志,1994,13(4):194-196. 被引量:6
  • 2Kawahito S, et al. Dynamic QRS-complex and ST-segment monitoring by continuous vectorcardiography during carotid endarterectomy. Br J Anaesth, 2003, 90:142-147
  • 3Micheli A, Medrano GA. Rational electrovectorcardiography. Arch Cardiol Mex,2001, 71:330-334
  • 4Riera AR, et al. Vector electrocardiographic alterations after percutaneous septal ablation in obstructive hypertrophic cardiomyopathy. Possible anatomic causes. Arq Bras Cardiol, 2002,79:466-475
  • 5Watanabe Y, et al. Vectorcardiographic evaluation of myocardial infarct size: depar ture parameters are superior to conventional spatial parameters. Jpn Circ J, 1998,62:473-478
  • 6Jensen J, et al. On-line vectorcardiography during elective coronary angioplasty indicates procedure-related myocardial infarction. Coron Artery Disease, 2000,11:161 - 169
  • 7Sawada K, et al. Spatial ventricular gradient in patients with Wolff-Parkinson-White syndrome in comparison with normal subjects: vectorcardiographic evidence for significant repolarization changes due to preexcitation. Intern Med, 1995,34:738-743
  • 8Davis LM, et al. Localisation of ventricular tachycardia substrates by analysis of the surface QRS recorded during ventricular tachycardia. Int J Cardiol, 1995,50:131 - 142
  • 9Chia JM, et al. Performance of QRS detection for cardiac magnetic resonance imaging with a novel vectorcardiographic triggering method. J Magn Reson Imaging,2000,12:678-688
  • 10Astrom M, et al. Vectorcardiographic loop alignment and the measurement of mor phologic beat-beat variability in noisy signals. IEEE Trans Biomed Eng, 2000,47:497-506

共引文献5

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部