三维超声重建测量肾脏体积的实验研究 被引量:11

Evaluation of renal volume with three dimensional ultrasonography an experimental study
下载PDF
导出
摘要 目的:评价三维超声对肾脏体积测量的准确性。方法:利用三维及二维超声经皮测量21枚活体肾脏体积(新西兰大白兔2只,成年杂种犬5只,小型香猪2头,拟切除移植肾3例)。移植肾和动物的肾剖腹后摘除通过水置换法测量体积。结果:三维超声和二维超声测量肾脏体积的准确率分别为97.36%和93.87%,与实际体积相比,三维超声测量值(r=0.9948,P=0.000)较二维超声测量值(r=0.9610,P=0.000)的一致性好。三维超声和二维超声测量的系统偏倚分别为0.75ml和1.67ml,一致性范围分别为-3.66~5.11ml和-8.15~10.79ml。结论:三维超声对肾脏体积的测量是可行的,具有较高的准确性和一致性。 Objective To evaluate the accuracy of three dimensional ultrasonography3DUS in the estimation of the kidney volume. Methods The volume of 21 kidneys was estimated with 3DUS and two dimensional ultrasound2DUS before they were removed from body2 rabbits 5 adult dogs 2 piglets and 3 patients. True volume of the kidney was determined by displacement of water in a beaker or measuring cylinder after the kidney were removed from the body. Results In vitro accuracy of 3DUS and 2DUS volumetery of kidney was 97.36% and 93.87% respectively. 3DUS was superior to 2DUS in term of assessing the volume of kidney and was characterized by a significantly smaller systematic bias and closer limit of agreement than 2DUS. Conclusion 3DUS volumetry measures kidney volume with high accuracy and was superior to 2DUS in kidney volume assessment.
机构地区 解放军总医院
出处 《中国临床医学影像杂志》 CAS 北大核心 2005年第2期86-88,共3页 Journal of China Clinic Medical Imaging
关键词 超声检查 多普勒 彩色 kidney ultrasonography Doppler color
  • 相关文献

参考文献5

  • 1姚绍球,王炼,罗岚,杨斌.B超测量移植肾体积及其临床意义[J].中华超声影像学杂志,1997,6(2):65-66. 被引量:11
  • 2徐辉雄,徐作峰,吕明德,谢晓燕,刘丽,刘广健,黄蓓.三维超声容积自动测量技术的准确性及重复性研究[J].中国超声医学杂志,2003,19(3):232-236. 被引量:50
  • 3Bland JM, Altman DG. Statistical methods assessing agreement between two methods of clinical measurement [J]. Lancet,1986, 1 (8476): 307-310.
  • 4Gilja OH, Thune N, Matre K, et al. In vitro evaluation of three-dimensional ultrasonography in volume estimation of abdominal organs[J]. Ultrasound Med Biol, 1994, 20 (2): 157-165.
  • 5Gilja OH, Smedvoll Al, Thune N, et al. In vivo comparision of 3D ultrasonography and magnetic resonance imaging in volume estimation of human kidneys [J]. Ultrasound Med Biol, 1995,21 (1): 25-32.

二级参考文献12

  • 1[1]Liang RI, Chang FM, Yao BL, et al. Predicting birth weight by fetal upper-arm volume with use of three-dimensional ultrasonography. Am J Obstet Gynecol, 1997,177:632-638
  • 2[2]Hsieh YY, Chang CC, Lee CC, et al. Fetal renal volume assessment by three-dimensional ultrasonography. Am J Obstet Gynecol, 2000, 182: 377-379
  • 3[3]Riccabona M, Nelson TR, Pretorius DH, et al.Distance and volume measurement using three-dimensional ultrasonography. J Ultrasound Med, 1995,14:881-886
  • 4[4]Xu HX, Zhang QP,Lu MD, et al. Comparison of two-dimensional and three-dimensional sonography in evaluating fetal malformations. J Clin Ultrasound, 2002, 30:515-525
  • 5[5]Xu HX, Lu MD, Zhou YQ, et al. Three-dimensional gray scale volume rendering of the liver: preliminary clinical experience. J Ultrasound Med, 2002, 21:961-970
  • 6[6]Xu HX, Yin XY, Lu MD, et al. Estimation of liver tumor volume using a three-dimensional ultrasound volumetric system. Ultrasound Med Biol, (in press)
  • 7[7]Xu HX, Yin XY, Lu MD, et al. Comparison of three- and two-dimensional sonography in diagnosis of gallbladder diseases: preliminary experience. J Ultrasound Med, (in press).2003,22:181~191
  • 8[8]Bland JM, Altman DG. Statistical methods assessing agreement between two methods of clinical measurement. Lancet 1986; 1(8476):307-310
  • 9[9]Riccabona M, Nelson TR, Pretorius DH, et al. In vivo three-dimensional sonographic measurement of organ volume: validation in the urinary bladder.J Ultrasound Med, 1996,15:627-632
  • 10[10]Hamper UM, Trapanotto V, Robert DeJong M, et al. Three-dimensional US of the prostate: early experience. Radiology, 1999, 212:719-723

共引文献58

同被引文献125

引证文献11

二级引证文献43

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部