期刊文献+

Sensitivity of a new-developed neutron detector

Sensitivity of a new-developed neutron detector
下载PDF
导出
摘要 We develop a kind of neutron detector, which consists of a polyethylene thin film and two PIN semicon- ductors connected face-to-face. The detector is insensitive to γ-rays. Its sensitivity to neutron has been calculated with MCNP program and calibrated by experiments, and the results indicate that the neutron sensitivity of the compensa- tion detector will vary with polyethylene converter. The compensation PIN detector can be employed to measure pulse neutron in neutron and gamma mixture radiation field. We develop a kind of neutron detector, which consists of a polyethylene thin film and two PIN semicon- ductors connected face-to-face. The detector is insensitive to γ-rays. Its sensitivity to neutron has been calculated with MCNP program and calibrated by experiments, and the results indicate that the neutron sensitivity of the compensa- tion detector will vary with polyethylene converter. The compensation PIN detector can be employed to measure pulse neutron in neutron and gamma mixture radiation field.
出处 《Nuclear Science and Techniques》 SCIE CAS CSCD 2005年第1期40-42,共3页 核技术(英文)
关键词 灵敏度 中子探测器 脉冲辐射 聚乙烯转炉 Sensitivity, PIN detector, Neutron, Pulse radiation, Polyethylene converter
  • 相关文献

参考文献3

  • 1MA Chi, Nucl Electron Detect Technol (in Chinese). 1995.15(6): 373.
  • 2Schutz R, Fehrenbacher A, Biersack J Pet al. Radiat Effects and Defects in Solids, 2002, 157:427-444.
  • 3LIU Qing-Zhao. Tile diagnosis technology of pulse radiation field (in Chinese). Beijing: Science Press, 1994:117.