摘要
目的:维修报警而无提示的简单故障和打印机不走纸的故障。方法:从分析故障原因入手,采用简单工具维修。结果:由供电系统造成的电容器击穿可更换电容器,使用中造成的打印机轴松动故障可复原固定。结论:紧急情况下及时处理仪器运行中出现的问题,使之始终处于最佳状态是检验工作质量的保证。
Reparing the trouble of the siren of the AVL995HB bolld gas analyzer which the siren does not shoe and the trouble of the printer which its paper does not move. Methods: Analyse the reason of the troubles, select the simlpe tools to repair them. Results. When the condenser is broken down by the power supply system. We can relace the condenser, when the printer's axle is loose during using, we may restore it and fix it. Conclusion: It is the quality guarantee of the work of inspection that we settle the problems of the working machine and it work in a good condition in time of emergency.
出处
《基层医学论坛》
2003年第3期267-267,共1页
The Medical Forum