F-820型血液分析仪白细胞分类结果评价

The evaluation of the results of leukocyte sort by F - 820 hemacytometer
下载PDF
导出
摘要 目的:为了考察F-820型血液分析仪白细胞分类结果的可靠性。方法:对173例随机病例分别进行自动和手工分类,其结果进行比较。结果:102例无显著性差异(P>0.05),60例有显著性差异(P<0.01),11例必须进行手工分类。结论:F-820型血液分析仪白细胞分类只能作为初筛方法,提倡仪器分类和镜检联合应用。 Objective: In order to evaluate the reliability of the results of the leukocyte sort which processed by the F - 820 hemacytometer. Methods: Compared with these two group results, the leukocyte of the 173 samples randomly was sorted by the F - 820 hemacytometer and the manmade method respectively. Results: The result of 102 samples showed not remarkable difference ( P>0.05 ) but 60 samples did ( P<0.01 ) .Another 11 samples must be sorted by manmade methods . Conclusion: The F - 820 hemacytometer was the screening test for the leukocyte sort. Advocate to use this instrument companioned by microscopic examination.
出处 《张家口医学院学报》 2001年第5期57-57,共1页