期刊文献+

空气细菌采样器室内采样效果比较 被引量:5

COMPARISON OF INDOOR COLLECTING EFFICACY OF AIR BACTERIAL SAMPLERS
下载PDF
导出
摘要 对不同工作原理的4种采样器进行空气中细菌采样效果比较。结果表明,安德森采样器对空气细菌的捕获率高,重复性好,并可测定细菌气溶胶粒子的大小,建议作为标准采样器使用。 The efficacy of four types of air sampler in sampling bacterial aerosol was compared in laboratory. The samplers compared were Andersen air sampler (made in USA), LCW-1 sampler (slit type, made in China), JWL-1 sampler (centrifugal type, made in China) and MF-45 sampler (millipore membrane type, made in China). The Andersen air sampler collected more bacterial particles than the other three samplers did. The number of bacterial particles collected by the latter three samplers was only 53~76% of that collected by the former. The average count median diameter of the bacterial aerosol measured by the Andersen air sampler was 4.3±1.1 μm.
出处 《中国消毒学杂志》 CAS 北大核心 1993年第1期26-29,共4页 Chinese Journal of Disinfection
关键词 空气 细菌 采样器 气溶胶粒子 air bacteria air sampler aerosol particle size