摘要
用CS-9000薄层扫描仪同时测定了安定、安眠酮及舒乐安定的血清浓度。采用正已烷-氯仿-甲醇(5:3:1)为展开剂,Rf值分别为0.35,0.64,0.29.3种药物测定波长分别为:254,249,264nm,可用于中毒抢救时3种药物的定性定量。
Simultaneously diazepam, methaqualoneand estazolam in human serum were deter-mined. Plate: silica gel GF254, mobile phase: n-hexane: chloroform: methyl (5: 3: 1),R_f value was: 0.53, 0.64, 0.29. Quantitativeassay was made by CS-9000 thin-layer-chro-matography flying-spot scanner with detectionat 254, 249, 264nm. Correlate coefficient was0.9968~0.9995.The rate of recovery was93%~98%. The precision of within plate andintraplate RSD was less than 3.7%. Themethod is simple, rapid, sensitive, and can beused in analysis of intoxication in emergencytreatment.
出处
《中国药学杂志》
CAS
CSCD
北大核心
1993年第9期550-552,共3页
Chinese Pharmaceutical Journal