期刊文献+

无创伤脉搏血氧测定仪的研制 被引量:5

The development of a noninvasive pulse oximeter
下载PDF
导出
摘要 本文给出了一种用于无创伤,连续地测量动脉血氧饱和度(SaO_2)的脉搏式血氧测定仪的原理、设计和仪器。脉搏血氧测定仪中使用的两个波长光是典型的红光和近红外光谱区的,正是对两波长相对光强比率的测量导致了血氧饱和度的计算。该脉搏血氧计采用了内含10位A/D转换器的Intel 8098单片机作为数据处理、计算和光驱动的控制。最后给出了初步的实验结果。 The theory, design and instrumentation of a pulse oximeter for the noninvasive and continuous measurement of arterial hemoglobin oxygen saturation (SaO_2) are presented. The two wavelengths of light used in the pulse oximeter are typically in the red and infrared ranges. It is the measurement of the relative intensities at these wavelengths that leads to the SaO_2 calculation. We have developed the oximeter using Intel 8098 single-chipcomputer with 10-bit A/D converter for the signal processing, calculation and light driver control. Finally, preliminary results of this oximeter are presented.
机构地区 华中理工大学
出处 《中国医疗器械杂志》 CAS 1993年第1期11-14,50,共5页 Chinese Journal of Medical Instrumentation
  • 相关文献

参考文献2

  • 1Setsuo Takatani,Peter W. Cheung,Edward A. Ernst. A noninvasive tissue reflectance oximeter[J] 1980,Annals of Biomedical Engineering(1):1~15
  • 2I. Yoshiya,Y. Shimada,K. Tanaka. Spectrophotometric monitoring of arterial oxygen saturation in the fingertip[J] 1980,Medical & Biological Engineering & Computing(1):27~32

同被引文献15

  • 1王海燕.肾病综合征时的急性肾功能衰竭[J].中华内科杂志,1995,34(3):149-149. 被引量:12
  • 2王质刚.血液透析与生存质量[J].中华内科杂志,1995,34(3):147-148. 被引量:25
  • 3张道友.血液透析患者营养不良研究近况[J].中华肾脏病杂志,1995,11(1):53-55. 被引量:18
  • 4尹良红,陈玉琼,林日英,林利雅,陈湛华,郑绮宜.应用心导纳图求积法原理检测透析病人血透前后心功能改变[J].中华肾脏病杂志,1995,11(3):166-167. 被引量:14
  • 5C. G. Olthof,P. M. Kouw,Ab J. M. Donker,J. J. Lange,P. M. J. M. Vries. Non-invasive conductivity technique to detect changes in haematocrit:in vitro validation[J] 1994,Medical & Biological Engineering & Computing(5):495~500
  • 6K. Yamakoshi PhD,S. Tanaka,H. Shimazu. Electrical admittance cuff for non-invasive and simultaneous measurement of haematocrit, arterial pressure and elasticity using volume-oscillometric method[J] 1994,Medical & Biological Engineering & Computing(1):S99~S107
  • 7J. P. P. M. Vries,C. G. Olthof,V. Visser,P. M. Kouw,P. M. J. M. Vries. Continuous measurement of blood volume using light reflection: Method and validation[J] 1993,Medical & Biological Engineering & Computing(4):412~415
  • 8Dr H. G. Goovaerts,F. R. Vries,J. H. Meijer,P. M. J. M. Vries,A. J. M. Donker,H. Schneider. Microprocessor-based system for measurement of electrical impedances during haemodialysis and in postoperative care[J] 1988,Medical & Biological Engineering & Computing(1):75~80
  • 9Dr. J. S. Wilkinson,S. J. Fleming,R. N. Greenwood,W. R. Cattell,C. Aldridge. Continuous measurement of blood hydration during ultrafiltration using optical methods[J] 1987,Medical & Biological Engineering & Computing(3):317~323
  • 10K. Yamakoshi,H. Shimazu,M. Shibata,A. Kamiya. New oscillometric method for indirect measurement of systolic and mean arterial pressure in the human finger. Part 1: Model experiment[J] 1982,Medical & Biological Engineering & Computing(3):307~313