SOMATOM DRH CT机US0080故障分析与排除

Analysis and Elimination of US0080 Error Information in SOMATOM DRH CT Set
下载PDF
导出
摘要 CT机在正常扫描情况下,图象上出现了一簇斜方向伪影。同时屏显US0080错误信息。从提示的错误信息分析,会有许多原因导致US0080信息出现.本文讨论了由于GCP信号瞬间丢失而产生的图象伪影。以及修复过程同时也找出了,由于BSP内5V电源低于正常值而产生的图象上有点状伪影的故障原因。 During CT Scanning there was a normal image in the direction of CCW.However in direction of CW the obliquestrenks appeared and US0080 mistake was showed on the screan.Author found:when the gunery rotated in the direction of CW.GCP single lamp go out for a moment,because GCP single line broke in a twinkling.Aften the line was changed the scanning progress recovered,but there were some white dotted artifacts around the image when SV power in BSP was increased from 4.93 V to 5.03 V,actted artifacts mentioned above disapeared.
作者 张延式
机构地区 广州军区总医院
出处 《CT理论与应用研究(中英文)》 1994年第2期40-41,共2页 Computerized Tomography Theory and Applications
关键词 伪影 GCP信号 CT 故障分析 排除 医用 s:Elimination, Error information