摘要
高效液相色谱测定血清皮质醇的方法,0.5ml血清加150ng地塞美松(内标剂)在55℃孵育15分钟后用5ml乙醚提取类固醇。移有机相于一尖底管、蒸干,残留物用100μl动相(1,2─二氯乙烷、异丙醇、环已烷按6/11/83容积混合,每升加冰醋酸0.2ml)重溶,注20μl于色谱系统。分析柱为Shim─packCLC─CN柱,柱温40℃,流速2.0ml/分。检测器波长242nm,0.001A满刻度。皮质醇和内标剂的相对保留时间分别为4.68和5.14分。分析回收率平均在98.1~100.0%之间,批内和批间CV<4%。本法在68.89~1103.54nmol/L之间呈线性,回归方程y=0.0242+0.0052x,r=0.9996,最低检出限13.8nmol/L。158名健康成人上午8~9时血清用本法测得皮质醇浓度为193.12土65.00nmol/L(70.07+23.56μl/L),男性(n=81)平均213.69土61.3Onmol/L(77.31±22.22μg/L)明显高于女性的(n=77)172.23±62.43nmol/L(62.46±22.63μg/L),P<0.001。
In
the present article,a metbod determining serum cortisol by highperformance liquid
chromatography was described.After 0.5ml of serum together with150ng of internal
standard,dexamethasone,was incubated in a 55℃ water both,steroids were extracted by
diethyl ether and the extract was evaporated.The residuewas redissolved with 100μL mobile
phase which consisted of 1,2─dichloroethane/iso─propanol/cyclohexane(6/11/83,by Vol.)
containing,per litre,0.2mL of gliacial aceticand then injected 20μL into chromatograp.Steroids
were separated by Shim─packCLC─CN column at a flow rate of 2.0mL/min and the eluates
were detected at 242nm.The relative retention time of cortisol and dexamethasone was 4.68
and 5.14min,respectively,The average analytical recoveries were between 98.1%and
100.0%.Thewithim─run CVs were 1.1%─4.0%and between─run was 3.8%.The linearity of
thismethod ranged from 68.89 to 1103.54nmol/L,with the lower detection limit of
13.8nmol/L(signal/noise=3).Serum cortisol concentration determined by this method in158
health adults was 193.12士65.00nmol/L,whose blood samples were obtainedbetween 8 to 9 in
the morning.The value in males(213.29土61.3nmol/L,n=81)wassignificantly higher(P<.001)than
that in femals(172.2土62.4nmol/L,n=77).
出处
《承德医学院学报》
1994年第3期172-176,共5页
Journal of Chengde Medical University
关键词
皮质醇
高效液相色谱
参考值
血清诊断
cortisol
high performance liquid
chromatography
reference value