摘要
Pb(Ⅱ)在0.2%乙二胺-5×10^(-3)mol·L^(-1)HOx-0.3mol·L^(-1)KOH溶液中,有良好的倒数示波图,其峰电位Ep=-1.05(vs.SCE),峰高与Pb^(2+)浓度在4.0×10^(-7)~2.0×10^(-5)mol·L^(-1)范围内成正比,检测下限可达2.0×10^(-7)mol·L^(-1)。本实验采用倒数示波计时电位法对铝合金"A"中铅进行了测定,并直接通过示波图,对络合物的吸附性进行了验证,获得了满意的结果。
In 2% ethylenediamine-5×10 ̄(-3)mol/L HO_x-0. 3 mol/L KOH(pH =-13.5) supporting electrolytes, a good oscillogram of pb(Ⅱ)and HO_x is obtained by reciprocaloscillographic chronopotentiometery. The peak potential is-1.05(vs. SCE ). The peak height is directly proportional to the concentration of pb(Ⅱ)in the range of 4.0×10 ̄(-7)~2.0×10 ̄(-5)mol/L. The detection limit iS 2.0×10 ̄(-7)mol/L. The method has been applied to determination of lead in allumniun alloy'A',The characteristics of the polarographic wave was tested and verified by oscillogram.The results are satisfactory.
出处
《分析测试学报》
CAS
CSCD
1994年第2期54-57,共4页
Journal of Instrumental Analysis
关键词
吸附络合物
示波计时电位
铅
溶液
Adsorbable complex, Reciprocal oscillographic chronopotentiometery, Lead determination.