摘要
本文对铟氢化物发生方法进行了探讨,所采用的方法使铟氢化物的发生效率明显提高。在电加热石英管原子化器中得到的测定灵敏度为0.007μg(1%吸收),比文献报道值提高了一个数量级。并讨论了铟氢化物的发生机理与原子化机理。
A method of indium hydride generation is modified and the results show that the efficiency of generation is increased significantly. The sensitivity of determination in the electrically heated quartz tube atomizer is 0. 007 μg/0. 0044 which is one order of magnitude better than that reported earlier. The mechanisms of generation and atomization in the quartz tube are discussed.
出处
《分析化学》
SCIE
EI
CAS
CSCD
北大核心
1994年第1期19-22,共4页
Chinese Journal of Analytical Chemistry
基金
国家自然科学基金资助课题
关键词
铟氢化物
原子化器
原子吸收光谱
Indium hydride, Electrically heated quartz tube atomizer, Atomic absorption spectrometry