摘要
测定了101例甲状腺病患者3、24h131I摄取率和30min9mTc摄取率,还测定了22例甲状腺正常者30min99mTc摄取率.结果是甲状腺正常的30min摄99mTc率为3.73±1.56;摄99mTc率在正常值以上的患者,其摄99mTc率与摄131I率的临床符合率为82.4%.在正常值以下的患者,其摄99mTc率与摄131I率的临床符合率为91.3%.在正常值范围的患者其摄99mTc率与摄131I率的临床符合率为88.5%。研究表明.摄99mTc率诊断甲状腺功能比摄131I率更方便。
The 3,24 h 131I and 30 min 99mTc uptake rates have been measured in 101 cases with thyroid disease. The 30 min 99mTc uptake rate was also measured in 22 cases with normal thyroid.The results showed that 30 min 99mTc uptake rate of normal thyroid were 3.73±1.56.The coincident rate of increased value in 99mTc uptake with fall uptake were 82.4%,while those of decreased and normal value were 91.3% and 88.5% respectively.Above studies have shown that in functional image of thyroid 99mTc uptake is more useful than 24h 131I uptake test.
出处
《核技术》
EI
CAS
CSCD
北大核心
1994年第5期269-271,共3页
Nuclear Techniques
关键词
甲状腺功能
试验
锝99M
^(99m)Tc
Thyroid uptake
Functional image
Coincident rate