摘要 对自制氰化高铁血红蛋白参考标准液进行了质量评价及初步应用。结果表明:此参考标准液的吸收曲线特征、A540/A504比例,波长750mm处吸光度、包装及无菌试验均达到ICSH标准;与上海医化所同类产品比较线性和准确度符合性好;4-8℃保存38月后其浓度最大波动为3.7%,平均2.5%。此标准液应用于581光电比色计及721、722分光光度计上作标准曲线以测定血红蛋白浓度,作为电子血细胞计数仪和血红蛋白仪测定血红蛋白时的校正液结果满意。 Quality evaluation studies and primary clinical uses of the HICN reference standard made in our laboratory are reported in this paper. Results prove that it is satisfactory in the characterostocs of absorbance curve,A540/A504 ratio, absorbance at 750nm, packing and bacterial culture test,all these parameters reaching the standards of ICSH,Comparing the linearity and accuracy with those of the products made in Shanghai Medical Laboratory Institute, the results of our products are in good agreement with them. Having stored for 38 months at 4~8℃ the HICN concentration only decreased 3.7%at most.2.5%in average. The reference standards have been used on 581,721,722 electrophotometers to make standard curve and on electron blood cell counter and hemoglobinometer as calibration solution for determination of Hb,results are satisfactory as well.
出处 《四川省卫生管理干部学院学报》 1994年第4期12-14,共3页 Journal of Sichuan Continuing Education College of Medical Sciences
关键词 氰化高铁 血红蛋白 参考标准度 血液检查 质量 HICN Reference standard quality eveluation uses