摘要 空气中磷化氢采集在内衬铝箔的复合膜袋中,取1ml注入气相色谱仪,经GDX-101柱分离后,火焰光度检测器检测。样品浓度为0.2~10.6mg/m3时,测定的RSD为4.0%~9.1%。样品在铝箔袋中保存,浓度可保持一周不变。本方法经辽宁省劳动卫生研究所验证,结果基本一致。 Air samples were collected in gas sampling bags made of aluminum and polyester and brought to the laboratory for analysis.1 ml of the sample was injected into a gas chromatograph equipped with a flame-photometric detector.A glass column (1m×3mm i.d.) packed with 40~60 mesh GDX 101 was used.Relative standard variations were in the range of 4.0%~9.1%for phosphine levels from 0.2~10.6mg/m3.Phosphine concentration was kept stable for a week when samples were stored in the bags. This method was verified by Liaoning provincial institute of occupational health.The results obtained were similar to the above.