期刊文献+

Zeeman-石墨炉原子吸收测定全血中微量铅 被引量:16

Determination of Trace Pb in Whole Blood by Zeeman-GFAAS
下载PDF
导出
摘要 提出用 0 . 5 %硝酸稀释样品 ,Zeeman 石墨炉原子吸收光谱法直接测定全血中微量铅 ,并讨论了有关的实验条件。方法的检出限为 10 0pg ,标样的相对标准偏差 (RSD) 0 . 36 %~ 1. 8% ,样品的相对标准偏差(RSD) 1. 4 %~ 2 . 3% ,回收率 96 . 0 %~ 10. 1 7% ,共存物干扰小。利用此法已分析 6 0 0多例铅接触工人和儿童 ,结果满意。本法简单 ,快速是一种理想的全血测定方法。 The determination method for trace Pb in whole blood was studied, which was diluted by 0.5% HNO3. In the measurement, Zeeman-GFAAS was used and the relative condition was discussed. This method shows that the detection limit was 10.0 pg, the relative standard deviation of standards (RSD) was 0.36%-1.8%, the relative standard deviation of samples (RSD) was 1.4%-2.3%. and the recovery rate was 96.0%-101.7%. The concurrent interference was low. This method is simple and rapid. It is an ideal method to determine trace Pb in whole blood.
出处 《光谱学与光谱分析》 SCIE EI CAS CSCD 北大核心 2005年第3期477-478,共2页 Spectroscopy and Spectral Analysis