摘要 ET -SPACETM全身热疗系统是利用红外线体表照射对人体进行全身加温的 ,可以确保人体各部位均匀加热。我科采用的治疗温度是 41 .8℃ ,升温时间约需 3h ,恒温期治疗时间为 1h ,降温时间约需 1h。其测温系统可以同时实时显示各路测温探测器的温度变化曲线 ;其控温系统对人体的绝对控温精度达到 0 . 1℃ ,显示分辨率达到 0 . 0 1℃。 The heating technique of infrared radiation to whole body surface is used by the ET-SPACE TM Whole Body Hyperthermia System,and this can make sure that each part of the body will be soaking heated The therapy temperature used by our department is 41 8 centigrade,and this temperature will be continued for 1 hour during therapy It will spend 3 hours to get to 41 8 centigrade from normal body temperature and from therapy temperature to safety body temperature is about 1 hour The measuring temperature system of ET-SPACE TM Whole Body Hyperthermia System can real-time show the temperature curve of each director at the same time,the absolute controlling temperature precision of it′s controlling temperature system is 0 1 centigrade,display resolution is 0 01 centigrade
出处 《医疗装备》 2005年第4期13-14,共2页 Medical Equipment
关键词 ET-SPACE^TM 全身热疗系统 红外线 体表照射 测温探测器 技术性能 whole body hyperthermia temperature measuring and controlling
  • 相关文献

参考文献5

  • 1Eisler K, Hipp R, Gpgler S, et al. New clinical aspects of whole body hyperthermia[ J ]. Adv Exp Med Biol, 1990,267: 393 - 8.
  • 2Westermann AM, Grosen EA, Katschinski DM, et al.A pilot study of whole body hyperthermia and carboplatin in platinumresistant ovarian cancer [J] .Eur J Cancer, 2001, 37 (9):1111-7.
  • 3Wiedemann GJ, Robins HI, Gutschs S, et al. Ifosfamide, carboplatin and Etoposide (ICE) combined with 41.8 degrees C whole body hyperthermia in Patients with refractory sarcoma [ J] .Eur J Cancer, 19996, 32A (5): 888 - 92.
  • 4Wiedemann GJ, Oleire F, Knop E, et al. Ifosfamide and carboplatin combined with 41.8 degress C whole body hyperthermia in Patients with refractory sarcoma and malignant teratoma [J] .Cancer Res, 1994, 54 (20): 5346- 50.
  • 5Wehner H, Von Ardenne A, Kaltofen S. Whole body hyperthermia with water filtered infrared radiation: technical physical aspects and clinical experiences [J] .Int J Hyperthermia, 2001,17 (1): 19-30.