摘要
本文根据瘦体组织电传导远比脂肪组织强的原理,研制了人体脂肪分析仪(BFA),采用4个电极,引出无痛信号(800μA、50kHz)进入人体,测试电阻抗(Z),编入计算机软件,运算人体脂肪量。选择30名18~20岁男性青年进行测量条件试验,得出食后半小时,上午、下午检测均可,喝少量水(少于500ml)对测量无明显影响(P>0.05)。另选52名受试者(男9,女43)年龄19~65岁,对用BFA测量的Z、体脂百分比(%BF)和用美国BC一300人体成分分析仪测量的电阻(R)、%BF进行了相关分析,r值分别为0.995(S(y.x))=7.97,P<0.01)和0.996(S(y.x))=0.61,P<0.01)。最后,用BFA对79名健康人(男22,女57)、年龄39.0±11.8岁、身高163±7.0cm、体重63.1±10.2kg测量体脂,男性体脂22.5±9.2%,女性体脂34.7±7.2%,用体脂判断肥胖比用体重判断肥胖发生率高,特别是女人、中老年人更明显。
body fat analyzer(BFA)involving tlie measurement of
bioelectrical impedance(Z)was described. It was based upon the principlc that the dectrical
conductivity of the lean body mass was far greater than thatof fat. Determinations of Z for the
estimation of fat with computer softwareWere made in 30 young men aged 18一20 y using an
electrical impedance ple一thysmograph with a four一electrode arrangement that introduced a
painlesssignal(800 μA at 50kHz) into the body. The measurements could be per-formed at
about 0.5h after breakfast or lunch,Drinking a small amount ofwater(<500 ml) gave no distinct
influcnce. The correlations between Z,percent body fat(%BF) by BFA and R,%BF by American
BC-300 BodyComposition Analyzer were r=0.995(S(y.x)=7.97,P<0.01),r=0.996(S(y.x))=0.67, P<0.
01) respectively, in 52 suUects(9 men, 43 women) aged 16一65y.For males and females,%BF
measured by BFA were 22.5±9.2, 34.7±7.2,rcspectively in 79 suUects(22 men and 57 wonien)
aged 39.0±11.8y, 163±7.0 cm height and 63.1 ± 10.2kg weight.The obesity rate diagnosed
by bodyfat was higher than by weight, more significatly for women and elderly.
出处
《营养学报》
CAS
CSCD
北大核心
1994年第1期78-82,共5页
Acta Nutrimenta Sinica
基金
北京军区科研课题
关键词
体脂肪
生物电阻抗
body
fat bioelectrical impedance