期刊文献+

一种消色差三磁铁偏转系统 被引量:4

A DOUBLY ACHROMATIC THREE-MAGNET BEAM DEFLECTION SYSTEM
下载PDF
导出
摘要 文章提出了一种新的消色差三磁铁偏转系统,并用矩阵法分析了系统的传输特性,导出了系统的消色差关系式。该系统的中心电子轨迹在束引出方向仅占据一个偏转半径的范围,比其他消色差系统部小.特别适用于医用电子直线加速器。 A new doubly achromatic three-magnet configuration is described which is used for low ener-gy electron beam deflection with largy energy spread.The transfer behaviour of the system isstudied by first-order transfer matrix method, and its achromatic relations are obtained.The ex-tent of central electron trajectory in elemetary dipole in the direction opposite to the exit trajectoryis only one bending radius.It is specially suitable for medical electron linac than all other achro-matic systems,for it may further minimizes the rotational radius of the gantry.
作者 王厚稳
机构地区 南京大学
出处 《原子能科学技术》 EI CAS CSCD 北大核心 1994年第4期336-340,共5页 Atomic Energy Science and Technology
关键词 医用加速器 磁铁系统 消色差 Medical electron linac Three-magnet deflection system boubly achromatic90°bending
  • 相关文献

参考文献1

  • 1陈银宝,原子能科学技术,1983年,5期,536页

同被引文献12

  • 1赖启基,王元林,文建平.高梯度行波电子直线加速器[J].物理学进展,1989,9(4):400-410. 被引量:2
  • 2Brown KL, Turnbull WE. Achromatic Magnetic Beam Deflection System[P]. US Patent: No3,867,635,1975.
  • 3Enge HA. Particle Accelerator Provided With an Adjustable 270° Non-dispersive Magnetic Charged-particle Beam Bender [P]. US Patent:No3, 379,911,1968.
  • 4Botman JIM, Bates T, Hagedoorn HL. A Doubly Focusing Magnet System for a Medical Linear Electron Accelerator[J]. Nucl Instrum Methods Phys Res B, 1985, 10/11:796~799.
  • 5王厚德,Chin J Nucl Phys,1995年,17卷,2期,179页
  • 6王厚德,核技术,1994年,17卷,1期,42页
  • 7钱云龙,核技术,1986年,10期,21页
  • 8陈银宝,原子能科学技术,1983年,5期,536页
  • 9赖启基,原子能科学技术,1986年,20卷,6期,686页
  • 10姚充国,原子核物理,1985年,7卷,2期,175页

引证文献4