摘要
将 ̄(10)B离子吸收室和 ̄(10)Be探测器在结构上合为一个整体,用一种气体一个入射窗有效地探测了AMS ̄(10)Be粒子。用Havar膜替代Mylar膜作入射窗,明显地降低了本底,相应的 ̄(10)Be/ ̄9Be原子比值达到约5×10(-13),将几个样品的 ̄(10)Be/Be测量值与已知值进行了比较。
The particles of AMS- ̄(10)Be are detected effectivelly with a detector into which  ̄(10)B ion cham-ber is incorporated using one gas and one entrance window. The Be background is reduced bymeans of Havar foil instead of Mylar as entrance window. The resulting background typically cor-responds to a  ̄(10)Be/ ̄9Be atomic ratio of about 5 × 10 ̄(-13) . The measured  ̄(10)Be/ ̄9Be ratios are comparedwith nominal values for tseveral samples.
出处
《原子能科学技术》
EI
CAS
CSCD
北大核心
1994年第6期551-555,共5页
Atomic Energy Science and Technology
基金
国家自然科学基金
关键词
加速器质谱计
探测器
本底
铍10
Accelerator mass spectrometry(AMS)Detector Background Atomic ratio