期刊文献+

不同长度微种植体支抗应力差异的三维有限元研究 被引量:45

Research on 3-D FEM stresses around anchorage micro-implants with different length
下载PDF
导出
摘要 目的 分析不同长度微型种植体在承载不同方向4 0 0g正畸力时的应力分布和大小。方法 利用三维有限元软件ANSYS建立不同长度的微型支抗种植体-骨有限元非线性分析模型,在加载侧方力(与种植体长轴成4 5°和90°)的条件下进行分析计算。结果 ①在两种加载条件下,不同长度的微型种植体应力值(包括单元应力和节点应力)呈规则分布,均显示8mm长度时,应力值最小,并且随着微种植体长度的增加或减小,应力值都呈增加趋势;②种植体颈部为应力集中区。结论 在承受非轴向力的作用下,8mm微种植体颈部的应力值最低,并且随微种植体长度的变化,应力都呈上升趋势。 Objective To investigate the stress around anchorage micro-implants with different length and select the appropriate micro-implant for orthodontic anchorage. Methods Three-dimensional element method (FEM) was used to analyze the stress around micro-implant with different length (5mm, 6mm, 7mm, 8mm, 9mm, 10mm), which was 400 grams loaded.The direction of the load was horizontal and 45 degree referring to the axis of micro-implant. Results ①The 8 mm micro-implant had mini-stress distribution and then the stress increased when the length of micro-implant changed;② the neck of micro-implant and the cortical bone were stress-focused areas. Conclusion The length can influence the stress around micro-implant. If possible, one should choose 8mm micro-implant or shorter.
出处 《口腔医学》 CAS 2005年第2期96-97,共2页 Stomatology
基金 江苏省卫生厅医学科技发展基金 (2 0 0 3 13 )
关键词 应力 长度 微种植体支抗 三维有限元 stress length micro-implant 3-D FEM
  • 相关文献

参考文献6

二级参考文献14

  • 1[2]Meijer HJ, Starman FJ, Boman F et al. Stress distribution around dental implants: influence of superstructure, length of implants, and height of mandible. J Prosthet Dent, 1992; 68(1): 96~ 100
  • 2[3]Lum LB. A biomechanical rational for the use of short implants. J. Oral Implant, 1991; 17(2): 126~ 131
  • 3[4]Lum LB, Osier JF. Load transfer from endosteal implants to supporting bone: an analysis using statics. Part two: Axial loading. J Oral Implantol, 1992; 18(4): 349~ 353
  • 4[5]Lum LB, Osier JF. Load transfer from endosteal implants to supporting bone: an analysis using statics. Part one: Horizontal loading. J Oral Implantol, 1992; 18(4):343~ 353
  • 5[6]Goodacre CJ, Kan JY, Rungcharassaeng K et al. Clinical complications of osseointegrated implants. J Prosthet Dent, 1999 May; 81(5): 537~ 552
  • 6[7]Grunder JH. A 3-year prospective multicenter follow-up report on the immediate and delayed-immediate placement of implants. Int J Ora Maxillofac-Implants. 1999 Mar-Apr; 14(2): 210~ 216
  • 7张少锋,马轩祥,欧阳官,徐君伍,刘宝林.种植全口义齿的应力分布特点[J].实用口腔医学杂志,1997,13(3):194-196. 被引量:18
  • 8张少锋,马轩祥,欧阳官,徐君伍,刘宝林.种植体长度和直径对种植全口义齿应力的影响[J].实用口腔医学杂志,1997,13(3):199-201. 被引量:41
  • 9梁星,赵志河,唐思青,白丁,吕丹,巢永烈,丁运萍.羟基磷灰石涂层钛种植支抗的稳定性[J].华西口腔医学杂志,1998,16(2):153-155. 被引量:13
  • 10梁星,唐思青,王虎,赵志河,巢永烈,吕丹,李满园,高卫民.羟基磷灰石涂层钛种植支抗牵拉磨牙的动物实验初探[J].华西口腔医学杂志,1998,16(2):156-157. 被引量:2