摘要 目的 介绍介入性MR引导下骨骼肌肉系统活检的技术操作方法,并评价其应用价值。方法 26例骨骼肌肉系统病变患者,均在介入性MR引导下进行经皮活检, 所用设备为开放式0 23T常导MR检查仪,并配有iPath200光学引导系统(Proview,Philips)。其中10例病人术后行外科手术治疗或活检。活检针为MR兼容性。结果 24例病人获得了组织病理学诊断( 24 /26 )。在10例后来行外科手术治疗或活检的病人中,MR活检诊断与最终诊断基本符合(10 /10)。未发现明显并发症。结论 介入性MR引导下骨骼肌肉系统病变的经皮活检是1种安全、准确的新技术。 Objective To introduce the procedures and evaluate the application value of the interventional MR-guided biopsy in musculoskeletal lesions. Methods Twenty-six patients with musculoskeletal lesions underwent interventional MR-guided percutaneous biopsies. All patients were performed under a 0.23 T C-arm configuration open MRI system (Proview, Philips Medical System) with optical tracking system (iPath 200, Philips Medical System). The needle was MR-compatible (Daum, Germany). 6 of 26 patients underwent MR contrast-enhanced examinations during the biopsies. The contrast agent used was gadopentetate dimeglumine (Gd-DTPA) 0.2 ml/kg body weight. 10 of 26 patients underwent a subsequent open surgical biopsy or surgical operative treatment. Results The histopathological diagnosis was got in 24 cases (24/26). In 10 patients who underwent subsequent open surgical biopsy or surgical operative treatment, primary diagnosis and final diagnosis was essentially consistent (10/10). No complications such as severe bleeding or neurological damage were seen. Conclusion Interventional MR-guided percutaneous biopsy in musculoskeletal lesions is an accurate and safe technique.
出处 《中华放射学杂志》 CAS CSCD 北大核心 2005年第5期493-496,共4页 Chinese Journal of Radiology
  • 相关文献

参考文献10

二级参考文献28

  • 1张雪哲,中华放射学杂志,1997年,31卷,375页
  • 2张雪哲,中华放射学杂志,1988年,22卷,136页
  • 3Mueller PR,Stark DD,Simenone JF,et al. MR-guide aspiration biopsy:needle design and clinical trial[J]. Radiology, 1986,161:605 - 609.
  • 4Steiner P, Botnar R, Dubno B, et al. Radio-frequency-induced thermoablation:monitoring with T1-weighted and proton-frequency-shift MR imaging in an interventional 0.5T environment [J]. Radiology, 1998,206:803-810.
  • 5Shankaranarayanan A, Duerk J L, Lewin J S. Developing a multichannel temperature probe for interventional MRI [ J ]. Magn Reson Imaging,1998,8:197-202.
  • 6Kochli V D, Mckinnon G C, Hofmann E, et al. Vascular interventions guided by ultrafast MR imaging: evaluation of differrent materials[J].Magn Reson Med,1994,31,3:309- 314.
  • 7Joensuu R P, Sepponen R E, Lamminen A E, et al. A shielded overhonser marker for MR tracking of interventional devices[J]. Magn Reson Med,2000,43:139- 145.
  • 8Wildermuth S,Debatain J F,Leung D A,et al. MR imaging-guided in travascular procedures: initial demonstration in a pig model[J]. Radiology, 1997,202:578 - 583.
  • 9Atalar E, Kraitchman D L, Carkhuff B, et al. Catheter-traking FOV MR fluoroscopy[J]. Magn Reson Med, 1998, 40: 865 - 872.
  • 10Bakker C J,Smits H F, Bos C, et al. MR-guided balloon angioplasty: in vitro demonstration of the potential of MRI for guiding, monitoring, evaluating endovascular interventions[J]. Magn Reson Imaging, 1998,8:245 - 250.

共引文献44

同被引文献39

引证文献4

二级引证文献11

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部