摘要
人工心脏瓣膜替换术后瓣周漏是一种严重并发症。我们报告2292例植入2575枚人工心脏瓣膜,发生瓣周漏者13例,发生率为0.51%(13/2575)。外院术后瓣周漏转入我院者5例。15例中再次手术,单纯修补5例,重新换瓣10例。本组死亡5例。另3例瓣周漏因心功能尚好,暂未手术。
Between May, 1976 and August,1991,2575 valves were replaced in 2292 patients.Periprosthetic leakage was found in 18 patients(including 5 performed at other hospitals)and occurred in prosthetic replacement of mitral valve in 11 cases and in aortic valve in 7. The incidence of this complication was 0.51%(13/2575).Reoperation for periprosthetic leakage was performed in 15 patients.Among these patients,5 had suture repair,and 10 had valve replacement.Hospital mortality from the reoperation was 33.3%(5/15).The predisposing factors of development of periprosthetic leakage include:(1)annular calcification or fragility;(2)annuloprosthetic disproportion;(3)excessive tension on sutures or annulus or both;(4)prosthetic infection;(5)inadequate suture material and size.
出处
《中华胸心血管外科杂志》
CSCD
北大核心
1994年第2期106-108,共3页
Chinese Journal of Thoracic and Cardiovascular Surgery
关键词
瓣周漏
人工心瓣膜术
并发症
Prosthetic replacement of cardiac valve
Periprosthetic leakage