术中双腔支气管插管位置监测方法的评估 被引量:3

Evaluation of the monitorings in positioning double-lumen endobronchial tubes during one-lung ventilation
下载PDF
导出
摘要 目的比较五种监测方法在术中核查双腔支气管插管位置变化的价值。方法选择需行单肺通气的手术患者60例,放置手术体位后同时监测气道阻力、脉搏血氧饱和度(SpO2)、呼气末二氧化碳分压(PETCO2)、流量-容量环和支气管气囊压力,3min后使用纤维支气管镜对双腔支气管插管位置进行判断及调整。结果气道阻力监测的灵敏度为63·64%,特异度为89·80%;PETCO2监测的灵敏度为54·55%,特异度为87·76%;流量-容量环监测的灵敏度为81·82%,特异度为85·71%。结论气道阻力监测、PETCO2监测、流量-容量环监测有助于临床判断双腔支气管插管位置。 Objective To evaluate five monitoring items in positioning the double-lumen tube(DLT) during one-lung ventilation. Methods Sixty selected patients undergoing one-lung ventilation were studied.Airway pressure, SpO_2, P_ETCO_2,flow-volume loops and bronchial cuff pressure were monitored continuously for three minutes after positioning the patients.Then fiberoptic bronchoscopy was used to check the position of DLT and correct it. Results The sensitivity for airway pressure monitoring was 63.64%, and specificity 89.80%, for P_ETCO_2 54.55% ,and 87.76%, for flow-volume loops 81.82%,and 85.71%. Conclusion During OLV,the monitorings of airway pressure, P_ETCO_2 and flow-volume loops are all helpful to find the displacement of DLT timely and effectively.
出处 《临床麻醉学杂志》 CAS CSCD 2005年第6期381-382,共2页 Journal of Clinical Anesthesiology
关键词 双腔支气管插管 气道阻力 呼吸旁气流 气囊压力 Double-lumen tube Airway pressure Side-flow analysis Positioning
  • 相关文献

参考文献8

  • 1Klein U, Karzai W, Bloos F, et al. Role of fiberoptic bronchoscopy in conjunction with the use of double-lumen tubes for thoracic anesthesia: a prospective study. Anesthesiology, 1998,88:346-350.
  • 2Lenoir RJ. Venous congestion of the neck;its relation to laryngeal mask cuff pressures. Br J Anaesth, 2004,93: 476-477.
  • 3Jonathan LB , David DA. Anesthesia for thoracic surgery. In:Miller RD,eds. Anesthesia. 5th ed. New York: Churchill livingstone, 2001. 1691-1700.
  • 4Szegedi LL,Bardoczky GI,Engelman EE,et al. Airway pressure changs during one-lung ventilation. Anesth Analg, 1997, 84:1034-1037.
  • 5Shafieha MJ, Sit J, Kartha R, et al. End-tidal CO2 analyzers in proper positioning of the double-lumen tubes. Anesthesiology,1986,64: 844-885.
  • 6Araki K,Nomura R, Urushibara R,et al. Bronchial cuff pressure change caused by left-sided double-lumen endobronchial tube displacement. Can J Anaesth, 2000,47: 775-779.
  • 7.诊断试验和筛检试验[A].见:陆培廉 主编.预防医学 第2版[C].人民卫生出版社,1995.337.
  • 8Schwarzkopf K, Schreiber T, Preussler NP, et al. Lung perfusion, shunt fraction, and oxygenation during one-lung ventilation in pigs: the effects of desflurane, isoflurane, and propofol.J Cardiothorac Vasc Anesth, 2003,17 : 73-75.

同被引文献17

引证文献3

二级引证文献7

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部