多元凹面换能器聚焦超声系统损伤离体牛肝的实验研究

EXPERIMENTAL STUDY OF ABLATION OF BOVINE LIVER TISSUE IN VITRO WITH MULTI-UNIT TRANSDUCER FOCUSED ULTRASOUND SYSTEM
下载PDF
导出
摘要 目的评价多元凹面换能器聚焦超声系统的有效性和安全性,为临床应用提供基础研究资料。方法在不同治疗剂量条件下,以该聚焦超声系统定点辅照离体牛肝,测定焦点温升及辅照后凝固性坏死体积,观察比较治疗前后靶区域超声灰度和病理变化。结果辅照后靶点温升可达80℃以上;随治疗剂量增大凝固性坏死长径和直径增大,体积增大;靶区超声灰度明显增强;光镜下见损伤区组织呈凝固性坏死,损伤区与正常组织分界清晰。结论该系统能有效地损伤靶点组织,使之发生凝固性坏死;并且辐照区域与周围组织分界清晰,诊断探头可实时监控靶点的灰度变化,具有良好的治疗安全性。 Objective To evaluate the efficacy and safety of Multi-unit Transducer Focused Ultrasound System. Methods Samples of bovine liver were ablated by various doses. Volume of coagulative necrosis, temperature rise of the focus, Ultrasound gray scale and pathological changes were recorded and compared. Results Temperature rise up to 80 ℃ can be acquired at the ablated area. Clear demarcation separating the necrosis and the normal tissues were observed. Conclusion The Multi-unit Transducer Focused Ultrasound System can effectively ablate the targeted area. Real time assessment of the gray scale changes of the targeted area can be acquired by the ultrasound probe.
出处 《实用临床医药杂志》 CAS 2005年第5期14-16,20,共4页 Journal of Clinical Medicine in Practice
基金 江苏省卫生厅重点资助项目(H200349)
关键词 聚焦超声 升温 生物学焦域 focused ultrasound temperature rise biological focal region
  • 相关文献

参考文献8

  • 1Kennedy J E, W u, F, ter Haar G R. High-intensity focused ultrasound for the treatment of liver tumours[J]. Ultrasonics, 2004 Apr, 42(1~9): 931.
  • 2张德俊.高强度聚焦超声换能器[J].中国超声诊断杂志,2000(2):1-4. 被引量:16
  • 3伍烽,王智彪,陈文直,白晋,邹建中,王芷龙,阮祥燕,杜永洪,孔繁兵,朱辉.高强度聚焦超声体外治疗肝肿瘤的剂量学研究[J].中华物理医学与康复杂志,2000,22(5):267-269. 被引量:22
  • 4Sibille A, Prat F, Chapelon J Y, et al. Extracorporeal ablation of liver tissue by high-intensity focused ultrasound[J]. Oncology, 1993, 50: 375.
  • 5Ter Haar G, Sinnett D, Rivens I. High intensity focused ultrasound-a surgical technique for the treatment of discrete liver tumours[J]. Phys Med Biol, 1989, 34(11): 1743.
  • 6Malcolm A L, ter Haar G R. Ablation of tissue volumes using high intensity focused ultrasound[J]. Ultrasound Med Biol, 1996, 22(5): 659.
  • 7Clarke R L, ter Haar G R. Temperature rise recorded during lesion formation by high-intensity focused ultrasound[J]. Ultrasound Med Biol, 1997, 23(2): 299.
  • 8Gelet A, Chapelon J Y, Bouvier R, et al. Local control of prostate cancer by transrectal high intensity focused ultrasound therapy: preliminary results[J]. J Urol, 1999 Jan, 161(1): 156.

二级参考文献1

共引文献36

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部