摘要
介绍了水泵瞬态试验装置和试验过程,对一台叶轮直径为125mm的管道离心泵进行了试验,通过快速响应测试仪表和高速数据采集系统实现瞬态参数的采集.给出在4种不同管路参数下离心泵启动过程的试验结果,并与基于准稳态假设的结果进行比较.试验结果表明,在启动过程中,离心泵的扬程随着转速的迅速上升而迅速增大,并在转速最大时达到最大值,转速稳定后的瞬态性能与稳态性能完全对应.在转速到达最大前的瞬态过程中,后阶段的性能与基于准稳态假设的结果基本吻合;在启动初始阶段,无量纲扬程呈现很大值,继而迅速减小到低于准稳态假设值,表现出明显的瞬态效应.
A volute centrifugal pump with impeller diameter of 125 mm was tested in a special facility designed and built for transient operation research. The transient parameters of centrifugal pump were collected by an acute instrument and a high-speed collection system. With the control valve held at different open positions during 4 different transient operations, the different transient results were analyzed compared with the assumed quasi-steady results. Analysis of the transient results shows that the flow-rate and pump head increase as rotational speed increases, and that they finally approach the steady-state values when rotational speed reaches maximal value. The transient performance after a moment accords with the performance under quasi-steady assumption. The nondimensional pump head coefficient tends to become larger than the quasi-steady value at the very beginning of the start, which indicates obvious transient effect.
出处
《浙江大学学报(工学版)》
EI
CAS
CSCD
北大核心
2005年第5期605-608,622,共5页
Journal of Zhejiang University:Engineering Science
基金
国家自然科学基金资助项目(50376056)
浙江省自然科学基金资助项目(M503073).
关键词
离心泵
快速启动
瞬态状况
管路阻力
Computational fluid dynamics
Computer simulation
Noise abatement
Pressure distribution
Pressure relief valves