ARCUS digma下颌运动轨迹记录仪获取架参数精确度的研究 被引量:9

Study on the accuracy of articulator parameters measured by the ARCUS digma system
下载PDF
导出
摘要 目的:以牙合架设定值为标准,研究采用ARCUSdigma下颌运动轨迹记录仪获取牙合架参数误差的大小。方法:将ARCUSdigma记录仪的测量装置固定在牙合架上,测量牙合架模拟下颌运动的参数。调整牙合架前伸髁导,从-30°到60°,每次间隔15°;Bennett角分别设定为4°、10°、20°、30°;前伸切导从0°到80°,间隔10°。每次调整设定值后测量5次,分析测量值和设定值间的误差。结果:前伸髁导测量值与设定值之差为-2.9°~2.4°,前伸切导的在-2.9°~2.9°,差异无显著意义(P>0.05)。Bennett角测量值均较设定值小,差值为-4.1°~-0.3°,有显著意义(P<0.01)。结论:采用ARCUSdigma下颌运动轨迹记录仪获取牙合架参数精确度高,有利于提高修复体咬合质量。 Objective: This study aimed to estimate the accuracy of articulato r parameters measured by the ARCUS digma system compared with preset values on a rticulator. Methods: The head frame was attached to the upper arm of the articul ator. The bite plate-transmitter assembly was mounted to the bottom part. The co ndylar settings of the articulator varied from -30 to 60 degrees in steps of 15 degrees. Likewise, the incisal guide settings varied from 0 to 80 degrees in st eps of 10 degrees. The Bennett angles were set at 4, 10, 20, 30 degrees respecti vely. The deviations between the preset values and those measured by the system were determined. Results: The errors of the condylar inclinations measured by th e system varied between -2.9~2.4 degrees, and those of the incisal guide angle w ere-2.9~2.9 degrees. There were no significant differences between the means and preset values (P>0.05). While the measured Bennett angles were mostly smaller t han preset values, the errors were -4.1^-0.3. There were significant difference between the means and preset values (P<0.01). Conclusions: With articulator para meters acquired by the ARCUS digma system, we can adopt the values of the condyl ar inclinations or the incisal guide angles for articulator adjustment. As to th e measured Bennett angle, it should be modulated.
出处 《口腔颌面修复学杂志》 2005年第2期86-88,共3页 Chinese Journal of Prosthodontics
关键词 he架 髁导斜度 切导斜度 Bennett角 ARCUS digma 下颌运动 轨迹记录仪 殆架参数 dental articulator inclination of condylar path inclination of inci sal guidance bennett angle
  • 相关文献

参考文献5

  • 1Pr(o)schel P, Thomas R. Momeburg, AfonsHugger, et al.Articulator-related registration-A simple concept for minimi zing eccentricocclusal error in the articulator[J]. Int J Prostho dont, 2002, 15: 289~294
  • 2Posselt U, Franzen G. Registration of the condylar path inclination by intraoral waxrecords: variations in three instruments[J]. J Prosthet Dent, 1960, 10 (3):441~454
  • 3李健慧,孙廉.用口内蜡记录法调节前伸髁导斜度的研究[J].中华口腔医学杂志,1991,26(2):90-92. 被引量:1
  • 4Curtis DA, Sorensen JA. Errors incurred in programming a fully adjustable articulatorwith a pantograph[J]. J Prosthet Dent, 1986, 55 (4):427~429
  • 5Celar AG, Tamaki K. Accuracy of recording horizontal cond ylar inclination and Bennettangle with the Cadiax compat[J].J Oral Rehabil, 2002, 29 (11): 1076~ 1081

二级参考文献2

  • 1洪流,中华口腔医学杂志,1985年,20卷,75页
  • 2叶秀芬,中华口腔医学杂志,1980年,15卷,198页

同被引文献92

引证文献9

二级引证文献25

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部