摘要 光学尘埃粒子计数器是利用尘埃粒子在强光照射下产生散射光,通过检测散射光强度设计而成,仪器接收灵敏度的稳定性是保证测量准确度的关键。提出了一种尘埃粒子计数器实现自动校正的方法,使仪器始终处于定标时的接收灵敏度。 A new approach was put forward to auto-correct the dust particle counter. The counter is used to measure the number and size of dust particles in clean air. The incident light is scattered by the dust particle and the scatter intensity is related to the size of the particle. So keeping constant detection sensitivity of the particle counter is the point for precise measurement. An auto-corrector was designed to adjust the sensitivity of BCJ-1 dust particle counter. The intensity of incident light and noise of the system are monitored. The auto-corrector can always retain the same sensitivity as the scaled sensitivity by compensation.
出处 《光学技术》 EI CAS CSCD 北大核心 2005年第3期413-414,419,共3页 Optical Technique
关键词 尘埃粒子 散射光 自动校正 dust particle scatter auto_correct
  • 相关文献

参考文献3

  • 1伏尔坚斯坦mB.分子光学(上册)[M].北京:高等教育出版社,1958..
  • 2黄惠杰,赵永凯,任冰强,程兆谷,杜龙龙,路敦武.尘埃粒子的半导体激光散射测量[J].中国激光,2002,29(12):1117-1121. 被引量:25
  • 3康华光.电子技术基础(模拟部分)[M].北京:高等教育出版社,2002..

二级参考文献13

  • 1J. Gebhart, P. Blankenberg, C. Roth. Counting efficiency and sizing characteristics of optical particle counters [C]. B. Y. H. Liu, D. Y. H. Pui, H. J.Fissan. Aerosols. New York: Elsevier, 1984. 7-10
  • 2W. C. Hinds, G. Kraske. Performance of PMS Model LAS-X optical particle counter [J]. J. Aerosol Sci.,1986, 17(1):67-72
  • 3R. G. Knollenberg. The measurement of latex particle sizes using scattering ratios in the Rayleigh scattering size range [J]. J. Aerosol Sci. , 1989, 20(3) :331-345
  • 4D. Bemer, J. F. Fabries, A. Renoux. Calculation of the theoretical response of an optical particle counter and its practical usefulness [J]. J. Aerosol Sci., 1990, 21(5):689-700
  • 5S. English. Detecting contamination in the semiconductor processing is key for particle counters [J]. Cleanrooms,1996, 10(11):25-32
  • 6Huijie Huang, Dunwu Lu, Lixin Zou et al.. A high-efficiency optical sensor for optical particle counters [C].SPIE, 1999, 3898: 382-387
  • 7A. Lieberman. Royco instruments particle counters:capabilities and limitations [C]. D. A. Lundgren, F. S.Harris, W. H. Marlow et al.. Aerosol Measurement,Gainesville: University Press of Florida, 1979. 183-193
  • 8B. Y. H. Liu, R. N. Berglund, J. K. Agarwal.Experimental studies of optical particle counters [J].Atmospheric Environment, 1974, 8:717-732
  • 9J. R. Hodkinson, J. R. Greenfield. Response calculations for light scattering aerosol counters and photometers [J].Appl. Opt., 1965, 4(11):1463-1474
  • 10C. F. Bohren, D. R. Huffman. Absorption and Scattering of Light by Small Particles [ M]. New York:John Wiley & Sons, 1983, Chapter 4. 82 - 129;Appendix A, 477-482

共引文献25

同被引文献27

引证文献4

二级引证文献9

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部