摘要
用免疫放射分析(IRMA)检测接受白细胞介素-2(rIL-2)治疗的肿瘤病人血清标本中IL-2活性。该法采用双抗体夹心模式,灵敏度<2.1U/ml,(290ng/ml),平均批内和批间变异系数分别为1.98%和5.27%,方法特异、灵敏、快速。用此法检测两组皮下注射不同剂量rIL-2的16例本院肿瘤病人血清浓度,并对部分病例进行药代动力学研究,其血药浓度变化曲线属二室模型。消除半衰期(t1/2β)为18h,达峰时间为2.5h。
Using two different monoclonal antibodies to human recombinant IL-2 (rIL-2), we developed a sandwish type immunoradiometric assay (IRMA) for quantitation of IL-2 in serum samples. This assay was more specific, sensitive and less time-consuming than other tests. The detection limit was below 2.1 U/ml (290ng/ml), the within-batch and between-batch CV were 1. 98% and 5. 27%, respectively. IL-2 levels in 16 serum samples from rIL-2-treated cancer patients were measured with present assay and its pharmacokinetics was analysed. The blood concentration change curve was of bicameral model and t1/2β was 18h and Tmax 2. 5h.