期刊文献+

计算机自动化X线头影测量分析系统的临床应用评价 被引量:2

Appraise to the Clinical Practice Situation of the Computerized Automatic Cephalometric Analysis System
下载PDF
导出
摘要 目的:对开发的计算机自动化X线头影测量分析系统(CACAS)的临床应用进行评价。方法:选择标准头颅侧位X线片16例,应用CACAS自动完成标志点定位,自动测量分析,获得各个标志点的坐标和测量参数值;以人机交互方式输入标志点,自动测量分析,获得各个标志点的坐标和测量参数值。结果:自动X线头影测量分析系统与计算机人机交互式两种头影测量方法间的比较,统计学上差异无显著性(除个别测量项目间)。结论:计算机自动化X线头影测量分析系统为更准确的测量分析提供了可靠、有效的工具。 Objective:Appraise to the clinical practive situation of the computerized automatic cephalometric analysis system(CACAS).Methods:Compared the parameter valuse of 16 skull lateral graphs analyzed in two ways.In one way,landmarks were automatically idetified by CACAS,then the coordinates and parameters were analyzed automatically.In the other way,landmarks' information was input by the way of human-computer interaction,then computer analyzed the coordinates and parameters.Results:There was no significant difference between most parameter values got in these two methods.Conclusion:CACAS may be a more relible and effective tool for cephalometry.
出处 《口腔颌面外科杂志》 CAS 2005年第2期152-155,共4页 Journal of Oral and Maxillofacial Surgery
基金 部科技基金资助项目(J99YG03)
  • 相关文献

参考文献6

  • 1Houston WJB, Maher RE, Mcelroy D. Sources of error in measurements from cephalometric radiographs [J]. Euro J Orthod, 1996,8(2):149-151.
  • 2Eppley BL, Sadove AM. Computerized digital enhancement in craniofacial cephalometric radiography[J]. J Oral Maxillofac Surg, 1991,49(4):1038-1041.
  • 3Macri V, Woniel A. Reliability of landmarks recording on firm and digital latual cephalograms [J]. Euro J Orthod,1993,15( 1 ):137-139.
  • 4Radolph DJ, Sinclair PG, Coggins JM, et al. Automatic computerized radiographic identification of cephalometric landmarks[J]. AM J Orthod Orthep, 1998,113 (2): 1-4.
  • 5魏明贵,周正炎,丁国伟,王国世,郭其一,杨杰.X线头影标志点的计算机自动识别初步研究[J].口腔颌面外科杂志,2000,10(4):283-287. 被引量:7
  • 6何艳,邵金陵,张辉,吴刚,吴莎.X线头影测量计算机自动识别系统的建立[J].中国美容医学,2004,13(1):62-64. 被引量:6

二级参考文献17

  • 1[1]Forsyth B,Shaw B,Richmond S,et al.Digital image of cephalometric radiography[J].Angle Orthod,1996;66(1):37.
  • 2[2]Jackson,Dickson,Birnie,et al.Digital image of cephalomerric radiographs:a preliminary report[J].Br J Orthod,1985;12:122.
  • 3[3]Kenneth R.Digital image processing[M].北京电子工业出版社(译),1998.
  • 4[4]Eppley MD,Michael MD.Computerized digital enhancement in craniofacial cephlometric radiography[J].J Oral Maxillofac Surg,1991;49:1038.
  • 5[5]Rudolph DJ,Sinclair PM,Coggins JM,et al.Automatic computerized radiographic identification of cephalometric landmarks[J].Am J Orthod Orthop,1998;113:1.
  • 6[6]Cohen AM,Linney AD.A priminary study of computer recognition and identification of skeletal landmarks as a new method of cephalometric analysis[J].Br-J-Orthod,1984;11:143.
  • 7[7]Parthasarathy S,Nugent ST,Gregson PG,et al.Automatic landmarking of cephalograms.Comput Biomed Res,1989;22:247.
  • 8[8]Tong W,Jensen GM,Fay DF,et al.Landmarking of cephalograms using a microcomputer system[J].Comput Biome Res,1990;23:358.
  • 9[9]Forsyth B,Davis DN.Assessment of an automated cephalometric analysis system[J].Euo J Orthod,1996;18:471.
  • 10Moyers RE,Bookstein FL.The inappropriateness of conventional cephalometrics[J].Am J Orthod,1979,75:599-617.