血液混样HCV/HIV-1 RNA全自动提取方法的应用评价 被引量:2

EVALUATION OF NUCLEIC ACID ENTIRELY AUTOMATED EXTRACTION METHOD ON HCV RNA/HIV-1 RNA IN BLOOD SCRENING
下载PDF
导出
摘要 目的:建立献血者全自动核酸检测方法,探讨在我国血液筛查中引进全自动核酸筛查的可行性。方法:选用STAR2000加样仪进行血样汇集(24人份),在COBASAMPLICOR进行扩增和检测分析结果,从灵敏度及抗干扰能力方面进行评价。结果:用国际标准核酸质控品考评及常规应用,表明全自动汇集、全自动核酸提取及扩增和检测95%的检出限量HCVRNA和HIV-1RNA分别为17·4IU/ml和20·6cp/ml,20mg/ml胆红素、1g/dl血色素及中等程度的脂血对结果无影响。通过对16512个样本共688汇集池分析,HCVRNA阳性1例(ELISA亦阳性),HIV-1RNA未检出阳性。结论:血液混样MPLC全自动核酸提取方法可应用于血液HCVRNA和HIV-1RNA的筛查工作。 Objective:To investigate and study Roche nucleic acid entirely automated extraction method on HCV RNA/HIV-1 RNA in blood sample pooled(24 samples) by sampling processor.Method:Entirely automated extraction method was used to extract nucleic acid simultaneously,the samples were amplicated and detected by Roche COBUS AMPLICOR system,the sensitivity,efficacy and anti-interference were evaluated.Result:The 95 percent cutoff HCV RNA/HIV-1 RNA tests by MPLC were 17.4 IU/ml and 20.6 cp/ml,20 mg/dl bilirubin,1 g/dl hemoglobin and medium lipid blood did not interfere with the results.1 of 16 512 donations was HCV RNA positive( positive for ELISA),none of HIV-1 RNA was detected.Conclusion:Entirely automated extraction method of HCV/HIV-1 RNA by MPLC can be applied in the blood screening.
机构地区 深圳市血液中心
出处 《现代预防医学》 CAS 北大核心 2005年第7期724-725,共2页 Modern Preventive Medicine
基金 瑞士罗氏分子诊断公司合作资助项目(20040195)
关键词 HCV RNA HIV-1 RNA 全自动核酸提取 评价 HCV RNA HIV-1 RNA Entirely automation Nucleic acid extraction Evaluation
  • 相关文献

参考文献6

  • 1叶萍,于卫建,胡荣花.核酸扩增检测技术预防输血传染病的应用现状及研究进展[J].国外医学(输血及血液学分册),2003,26(5):399-402. 被引量:9
  • 2Engelfriet, C.P, Reesink, H.W. Implementation of donor screening for infectious agents transmitted by blood by nucleic acid technology (International Forum) [J] . Vox Sanguinis, 2002, 82, 87-111.
  • 3T. M. Cuijpers. Validation of NucliSens Extractor in combination with the hepatitis C virus Combas Amplicor 2.0 assay in four laboratories in the Netherlands Utilizing nucleic acid amplification technology for blood screening [J] .Vox Sanguinis, 2001 (81), 12-20.
  • 4Meng, Q., Wong, C., Rangachari, A., et al. Automated multiplex assay system for simultaneous detection of hepatitis B virus DNA, hepatitis C virus RNA, and human immunodeficiency virus type 1 RNA [J]. Journal of Clinical Microbiology. 2001, 39, 2937- 2945.
  • 5Grant, P.R., Sims, C.M., Krieg-Schneider, et al. Automated screening of blood donations for Hepatitis C Virus RNA using the Qiagen BioRobot 9604 and the Roche COBAS HCV Amplicor assay [J] . Vox Sanguinis, 2002, 82, 169- 176.
  • 6Gabriele Holzl. Entirely automated Quantification of Human Immunodeficiency Virus Type 1 RNA in Plasma by Using the Ultrasensitive COBAS AMPLICOR HIV-1 Monitor Test and RNA Purification on the MagNA Pure LC Instrument [J] .Journal of Clinical Microbiology, 2003, 41,1248 - 1251.

二级参考文献26

  • 1李执如.国外医学:输血及血液学分册,2001,24(6):537-537.
  • 2Vargo J ,Smith K,Knott C,et al. Transfusion,2002,42(7):876.
  • 3Lelie PN, Drimmelen HA, Cuypers HT, et al. Transfusion. 2002.42(5) :527.
  • 4Mitsunaga S, Fujimura K, Matsumoto C, et al. Tranafuaion, 2002,42(1):100.
  • 5Coombs RW,Welles SL, Hooper C, et al. J Infect Dis, 1996,174(4) :704.
  • 6Grant PR ,Sims CM ,Tedder RS. Transfusion,2002,42(8) : 1032.
  • 7Nubling CM, Unger G, Chudy M, et al. Transfusion,2002,42(8) : 1037.
  • 8Widell A, Molnegren V, Pieksma F, et al. Transfus Meal, 2002,12 (2) :107.
  • 9Gallarda J L, Dragon E. Mol Diagn, 2000,5(1) : 11.
  • 10Lee HH,Allain JP. Vox Sang, 1998,74 (suppl 2) :119.

共引文献8

同被引文献28

引证文献2

二级引证文献14

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部