期刊文献+

UF-100全自动尿沉渣分析仪检测尿液红细胞的应用 被引量:3

Valuation of UF-100 Automatic Flow Urinary Sediment Analysis Bend in Red Blood Cell in Urine
下载PDF
导出
摘要 [目的]评价某公司的UF100全自动流式尿沉渣分析仪(下简称UF100)检测尿液中红细胞(RBC)的性能。[方法]按照NCCLSLiteratureGP16A和尿沉渣检查标准化的建议和要求,选择750份临床标本用上述的方法进行分析评价。[结果]UF100计数尿液中红细胞结果与传统的手工牛鲍氏计数板法有良好相关性,γ=0.973,t检验结果P>0.05;RBC<5000个/μl时,有良好的线性与准确性;在RBC>40个/μl时,CV<6%。尿液中的细菌、真菌、草酸钙结晶能对红细胞的测定造成干扰。[结论]UF100是一种尿液中红细胞定量分析快速筛查的方法,建议采用其作为常规的尿沉渣筛查工具。 [Objective]To valucate the performance of the method used in detecting red blood cells(RBC) in urine by Sysmex UF-100 automatic flow urine Analysis(UF-100).[Methods]Collecting 750 cases of clinical urine samples under the requirement of NCCLS Literature GP16-A,analysis by above method.[Results]It has a favorable relativity in detecting of urine RBC by UF-100 and Neubauer’s Counting Board.RBC γ=0.973,t examination(P>0.05);Good linearity in range of 5 000 RBC/μl and high degree.CV<6% when the number of RBC is samller than 40 RBC/μl.The bacteria and candida、have different effects in RBC detection in urine.[Conclusions]UF-100 is a fast and easy testing methond to perform urine sediments quantitative erinalysis.We sugest to choose the Sysmex UF-100 as sieving rule testing RBCs methond.
出处 《预防医学论坛》 2005年第4期421-422,426,共3页 Preventive Medicine Tribune
关键词 UF-100全自动式尿沉渣分析仪 尿红细胞 干扰 UF-100-full-auto Analyzer Urine RBC Effect
  • 相关文献

参考文献7

  • 1Winkel P,Statland BE,Jorgensen K. Urine microscopy,an illdefined method, examined by a multifactorial technique[J]. Clin Chem1974,20,436-439.
  • 2National Committee for Clinical Laboratory Standards. Urinalysis and collection and preservation of urine specimens;approved guideline. NCCLS document GP 16-A. Wayne, PA [ M ] . NCCLS,1995.21.
  • 3中国临床检验标准化专业委员会.尿液分析和尿液样品的收集、运输和储存:批准指南[M].北京:CCCLS,2001.22-26.
  • 4Ottiger, Huber. R, Quantitative urine Particle Analysis: Integrative approach for the optimal combination of Automation with UF-100and microscopic review with KOVA Cell chamber[J]. Clin Chem,2003,49,617-623.
  • 5金大鸣.UF-100型尿沉渣全自动检测仪评价[J].中华检验医学杂志,1998,22(3):61-62. 被引量:145
  • 6王娟 刘寓.草酸钙结晶对UF—100分析仪红细胞测定影响的探讨[J].中华医学丛刊,2003,3(10):21-22.
  • 7马骏龙,丛玉隆.菌尿对尿液分析仪测定尿红细胞的影响[J].中华医学检验杂志,1999,22(4):205-207. 被引量:145

二级参考文献1

共引文献272

同被引文献15

  • 1崔海霞.UF-100全自动尿沉渣分析仪检测尿红细胞假阳性结果分析[J].延边大学医学学报,2008,31(3):193-194. 被引量:1
  • 2罗燕飞,龚彩平,叶兆雄.UF-100全自动尿沉渣分析仪对红细胞、白细胞检测的影响因素分析[J].实用医学杂志,2007,23(15):2429-2430. 被引量:8
  • 3许文荣.临床检基础验学[M].北京:高等教育出版社,2006.155-156.
  • 4Li Peng.such as UF-100 fully automated urine analyzer flow of influencing factors[J].Journal of modern laboratory medicine,2007,(222):35-36.
  • 5Altekin E,Kadicesme O,Akan P,et al.New generation IQ-200automated urine microscopy analyzer compared with KOVA cellchamber[J].J Clin Lab Anal,2010,24(2):67-71.
  • 6Regeniter A,Haenni V,Risch L,et al.Urine analysis performed byflow cytometry:reference range determination and comparison tomorphological findings,dipstick chemistry and bacterial culture re-sults-a multicenter study[J].Clin Nephrol,2001,55(5):384-392.
  • 7Jiang T,Chen P,Ouyang J,et al.Urine particles analysis:perform-ance evaluation of Sysmex UF-1000i and comparison among urineflowcytometer,dipstick,and visual microscopic examination[J].Scand J Clin Lab Invest,2011,71(1):30-37.
  • 8Shayanfar N,Tobler U,von Eckardstein A,et al.Automated urinaly-sis:first experiences and a comparison between the Iris IQ200 urinemicroscopy system,the Sysmex UF-100 flow cytometer and manualmicroscopic particle counting[J].Clin Chem Lab Med,2007,45(9):1251-1256.
  • 9丛玉隆,马骏龙,李云莲,周建山,丁颖,刘琴风,陈淑云,岳秀玲,陈瑾.尿沉渣测定四种方法的参考值范围调查[J].中华医学检验杂志,1997,20(4):229-231. 被引量:137
  • 10范红平,朱星成,张群智.尿沉渣定量分析仪检测尿液红细胞的临床价值[J].检验医学与临床,2010,7(2):126-127. 被引量:13