摘要 可用于高效液相色谱(HPLC)电化学检测器的碳纤维电极,受流速影响小,在流通体系中响应快,重现性好,灵敏度高。以它作为工作电极的检测器,用于流动注射安培法时,对3,4一二羟基苯甲胺氢溴酸盐(DHBA),其线性范围为0.2~160ng,比一般流通体系检测器提高约3~4个数量级,灵敏度亦较常用的检测器高,检测限对于DHBA可达0.07ng。 New Carbon fiber electrodex hare recently hcem designed 、made and evaluated. Viscose fber elec-trode the best of its kind,electrode was used as working electrode in electrochemical detectorcombined with HPLC. Results evaluated show that the new electrode is independent of folwrate and has high sensibility in flow system Detector with viscose fiber electrode as the workingelectrode was mocd to determine 3,4-dihydroxy -benzyamine hydrobromide(DHBA)with flowinjection amperoetric detection,the response keing linear over the tset range of 0.2ng-160gn,detection limits for DHBA heing 0.07gn The sensibility of this electrode is higher by 3-4 or-der of magnitude than other detectors usually used in jlow system.
作者 冯德雄 方伦
机构地区 暨南大学化学系
出处 《暨南大学学报(自然科学与医学版)》 CAS CSCD 1994年第1期62-66,共5页 Journal of Jinan University(Natural Science & Medicine Edition)
基金 国务院侨办重点科学基金资助项目
关键词 电极 碳纤维 电化学检测器 electrode carbon fiber electrochemical detector high performance liquid chroma-tography