相差显微镜和UF-100联合检测鉴别血尿来源 被引量:23

Identification of hematuria by combination of phase contrast microscope of Sysmex UF-100 cytometer
下载PDF
导出
摘要 目的探讨相差显微镜和UF-100联合检测在鉴别肾性与非肾性血尿中的价值。方法用相差显微镜和UF-100联合检测105例不同来源血尿标本的尿RBC数、尿RBC形态及尿RBC平均前向闪射光强度(RBC-MFsc)等。结果70例肾性血尿尿RBC数通常<500/μl,尿RBC畸形率84%;当RBC-MFsc≤70ch(Channel)时,敏感性90.2%,特异性83.6%,诊断符合率92.8%。35例非肾性血尿的主要特点是:尿RBC数通常>500/μl,尿RBC畸形率14%;RBC-MFsc通常>80ch。结论联合检测分析尿RBC数、尿RBC形态及RBC-MFsc等指标能客观、准确地鉴别和诊断血尿来源。 Objective To evaluate the values of combined detection for distinguishing glomerular hematuria from nonglomerular hematuria by phase contrast microscope with Sysmex UF-100 automated flow cytometer.Methods The count and morphology of urinary red cell (RBC) in 105 urine samples from hematuria patients were measured by phase contrst microscope and urinary RBC-mean forward scattered light intensity (RBC-MFsc) were measured by Sysmex UF-100 automated flow cytometer simultaneously.Results In 70 patients with glomerular hematuria the number of urinary red cells was usually less than 500/μl and the percentage of deformed red cells was 84%.When the cut-off value of RBC-MFsc was defined as less than 70 ch for glomerular hematuria,the sensitivity was 90.2%, specificity was 83.6% and accuracy was 92.8%.In 35 patients with nonglomerular hematuria the number of urinary red cells was more than 500/μl,the percentage of deformed red cells was 14% and RBC-MFsc was more than 80 ch.Conclusion The combined examination including quantity,morphology and MFsc of urinary red cells may be able to disclose and diagnose the source of hematuria more objectively and accurately.
出处 《临床检验杂志》 CAS CSCD 北大核心 2005年第4期281-283,共3页 Chinese Journal of Clinical Laboratory Science
基金 江苏省高校自然科学基金(04KJD180045)
关键词 UF-100 相差显微镜 联合检测 血尿来源 鉴别 尿RBC 非肾性血尿 诊断符合率 检测分析 500 畸形率 尿标本 敏感性 特异性 形态 hematuria urinary red cell morphology RBC-mean forward scattered light intensity phase contrst microscope Sysmex UF-100 automated flow cytometer
  • 相关文献

参考文献6

二级参考文献14

  • 1韩俊宏,李影.用尿红细胞容积分布曲线鉴别诊断血尿的来源[J].陕西医学检验,1995,10(3):39-39. 被引量:1
  • 2熊立凡.临床检查基础[M](第三版)[M].北京:人民卫生出版社,2003.196.
  • 3[3]NCCLS. Urinanalysis and collection, transportation, and preservation of urine specimens; approved guideline [S]. NCCLS document GP 16 A. NCCLS, wayne, Pennsylvania, USA,1995,7~10.
  • 4[5]Kurt HP, Kampf Sc. Automation of urine sediment examination (acompanison of the sysmex UF-100 automated flow cytometer with routine manual diagnosis)[J]. Clin Chem, Lab Med 199,37(7):753~759.
  • 5[7]Langlosis MR, Delanghc JR, Steyaert SR, et al. Automated flow cytometry compared wigh an automated dipstick reader for urinalysis[J]. Clin Chem,1999,45(1):118~122.
  • 6[8]Kouri T, Kahkoner U. Malminiemik R et al. Evaluation of sysmex UF-100 flow cytometer vs chamber counting of supravitally stained specimens and conventional bacterial cultrues[J]. Am J Clin Path,1999,112:25~35.
  • 7[9]Koruri T, Fogazzi G, Gant V, et al. European urinalysis guideline scand[J]. J, Clin Lab Invest, Vol 60-supplement 2000: 231.
  • 8Kurt.HP,et al.Autonation of Urine Sediment Examination:a Comparison of the Sysmex UF-100 Automated Flow Cytometer with Routine Manual Diagnosis(Microscopy,Test strips,and Bacterial Culture)[J].Clin Chem Lab Med,1999;37(7):753~764.
  • 9Langlois,et al.Automated Flow Cytometry Compared with an Automated Dipstick Reader for Urinalysis[J].Clin Chem,1999;45(1):118~122.
  • 10NCCLS.Urinalysis and Collection,Transportation,and Preservation of Urine Specimens:Approved Guideline.NCCLS Document GP16A.NCCLS.Wayne Pennsylvania,USA,1995:7~10.

共引文献133

同被引文献104

引证文献23

二级引证文献93

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部